Product Main

Specifications

Double MPU to ensure system security
Manual bolus but automatical antibolus to increase patient safety
Automatically recognize the syringe type--10ml/20ml/30ml/50ml/60ml,any brand after calibration
Up to 8 hours battery back up for single channel at the rate of 5ml/h, 6 hours for double channel, 4 hours for four channel
Ideally suitable for OT, ICCU, NICU, PICU etc 
Specification 
Model
Parameters   
NC-FA343
Channel
Four
Compatible syringe
10ml 20ml 30ml 50/60ml
Infusion rate
0.1-400ml/h, (50ml/60ml)
0.1-200ml/h, (20ml/30ml)
0.1-99.9ml/h, (10ml)
Purge rate/bolus rate/top infusion rate
400ml/h;(50ml/60ml)
200ml/h;(20ml/30ml)
99.9ml/h;(10ml)
Volume limit
0-999.9ml,Step by 0.1
Accuracy
+2%
Infusion Step by
0.1ml/h
Visible & audible alarms
Syringe dislocation, Occlusion, Near empty, Pause overtime, Low battery, VTBI completion, Power cord falling off, Malfunction, Incorrect syringe type, Electrical circuit jam
Occlusion pressure: high
20ml\30ml\50ml/60ml Syringe Space: 90kpa+30kpa
                   10ml Syringe Space: 100kpa+30kpa
Occlusion pressure: low
20ml\30ml\50ml/60ml Syringe Space: 50kpa+30kpa
                 10ml Syringe Space: 70kpa+30kpa
KVO rate
0.1ml/h+0.02ml/h
Power
AC: 100v-240v 50HZ/60HZ  DC: 12V
Pause overtime alarm
1min50s-2min
Appalled standards
ISO 13485 & CE Certified
Dimension
425mm(L)*125mm(W)*245mm(H)
Weight