Single Channel ECG machine with measurement,Real-time ECG waveform,20 cases memory, Built-in USB
ECG-NC13 Interface (ECG software optional) DC/AC power with built-in rechargeable Ni-MH battery,320x240 graphic 3.8 inch LCD
Features
1 Design patent software patent
2 Application of CSE database, reliable measurement and analysis results
3 12-lead simultaneous acquisition
4 Detection and alarm on lead-off and low battery capacity
5 Automatic adjustment of baseline for optimal printing
6 Built-in ECG simulator for DEMO purpose
7 3 kinds of operation mode: AUTO. MAN and ANA
8 3.kinds of filter: HUM. EMG and ADS
9 AC, DC and battery power supply
10 2 types lead mode: Standard and Cabrera
11 15ECG files can be save in the ECG
12 250G files can be save to SD card
13 Communication with PC (Option)
Specification
Safety Standard:
|
IEC Class 1,type CF
|
Acquisition Mode:
|
12 lead acquisition simultaneously
|
Input Circuit:
|
Floating,protection against defibrillator and pacemaker
|
A/D converter:
|
12 bit
|
Time Constant:
|
≥ 3.2 s
|
Frequency Response:
|
0.05Hz~150Hz(-3dB)
|
Sensitivity :
|
2.5,5,10,20,40(mm/mV).Auto
|
Input Impedance:
|
≥50MΩ
|
Input Circuit Current:
|
≤50nA
|
Calibration Voltage:
|
1mV±2%
|
Noise Level:
|
<15µVp-p
|
Anti Baseline Drift :
|
Automatic
|
Filter :
|
EMG filter 35Hz~45Hz(-3dB)
ADS filter:Auto( 0.15~0.50Hz)
HUM filter:50Hz/60 Hz(-20dB)
|
CMRR:
|
>100dB(with AC filter)
|
Patient Current leakage:
|
<10µA
|
Paper Speed:
|
6.25/12.5/25/50 mm/s
|
Recording Paper:
|
50 mm(width),20 m/30 m(length),roll paper
|
Power Supply:
|
AC:110/220V,50/60Hz;85V-265V(option)
DC:12V,built-in rechargeable lithium battery
|
Dimension :
|
295mm*260mm*60mm,415mm*195mm*380mm(Carton)
|
Net Weight:
|
2.3kgs
|
Gross Weight:
|
4.3kgs
|
Standard Configuration:
Main Unit
Patient cable
Chest electrode
Limb electrode
Thermal paper
Paper roller
Rechargeable battery
Dust cover
Options:
PCEGG-500A Workstation
Software package
Aluminum trolley
USB function board