Product Main

Quick Details

Place of Origin: China (Mainland)
Brand Name: Poweam
Model Number: G6H
Screen Size: 8.4
Screen: color TFT LCD screen

Specifications

●Displayer:8.4" color TFT LCD screen with maximum 7waveform display 

●Adjusting of lightness continuously make patient more comfortable in midnight

Battery: Rechargeable high-energy built-in battery
Advanced streamline outline design, portable, compact, lightweight
Operating menu with multi-language interface selection: English, Spanish, Portuguese, Chinese,
French, Arabic, Russian, etc.

Easy operation with user friendly menu structure design and rotary dial
Against electrosurgical interference and defibrillation and no need to disconnect the monitor from 
the patient in process of defibrillating.

WAN communication function to network with central monitoring system and make long-distance monitoring, diagnosis, maintenance and software upgrade possible
Optional built-in wireless networking function
Support external printer to record graph and trend
Intelligent audio and visual comprehensive alarm

Suitable for use in physicians' offices, clinics, outpatient surgical centers, extended care facilities 
and other patient care areas, which of require affordable monitoring
Standard configuration: ECG, HR, PR, SpO2, ST analysis, RESP*2 (RA-LL impedance and nasal cavity), 
NIBP (with venous punch), TEMP*2(surface and rectal type), Drug dose calculation.

Optional configuration: Wireless connection, Recorder and, 2-IBP

When the parameter value activated alarm, they will be shown in 3 different colors. (Red, Yellow and White)

 

Technical Parameter

Temperature (surface and rectal)
Measurement range: 25.0 – 45.0°C 
Accuracy: ± 0.1 °C
Resolution: 0.1 °C
Response time: ≤ 3 min


ECG

Input: 5 wires ECG cable
Lead section:Ⅰ,Ⅱ,Ⅲ/ aVR, aVL, aVF/ V

Gain(mm/mV): 1/41/2124
Frequency response: Diagnosis  0.05 - 100Hz   Monitor 0.5 - 40 Hz   Surgery 1.0 - 25 Hz
Sweep speed(mm/sec): 12.5, 25, 50
Heart rate range: 30-254 BPM
Heart rate accuracy: ± 1 %
ST segment deviation analysis: -0.8-+0.8mV
ST accuracy: ±0.02mv

Respiration Rate(RA-LL impedance and nasal cavity)
Measurement range: 0 - 120 BPM (adult)
Accuracy: ± 1 BPM or 5 %
Resolution: 1 BPM

Environment TemperatureOperation 0.5 - 40°CStorage -20 - 50 °C
Relative humidity:≤ 80%            


Pulse Rate
SPO2 measurement range: 0 - 100 %
Resolution: 1 %
Pulse measurement range: 30- 254BPM
Accuracy: ± 2 %
Resolution: 1 BPM

 NIBP (Non-invasive blood pressure) 

 

Measurement type : adult, pediatric, neonatal
Measurement range: Systolic 4.0 - 37.0 kPa Diastolic  1.3 - 33.0 kPa Mean 2.6 - 35.0 kPa
Accuracy : ± 0.4 kPa or 5 %
Resolution: 0.1 kPa
Protection: over pressure


IBP(Invasive blood pressure) (with venous punch) 
(Option)

Measurement range: -1.3~40kPa (-10~300mmHg)
Channel: 2 channel
Transducer sensitivity: 5MV/V/mmHg
Unit display: KPa or mmHg selectable


Power requirements

Input:  98~260 V AC,  50/60Hz
Consumption: ≤ 80VA