Product Main

Specifications

Key Specification
 
 
Model
OTS-A05
Dimensions of the Working Chamber 
1000*1000*800 mm (W×H×D)
External Dimension of Chamber
1500*1860*1370MM  (W×H×D)
Temperature Range 
-50-200℃, can be set
Rate of Temperature Increase
 
Room temperature----->150℃    about 40minutes(Nonlinear no-load, about 3.5℃/min)
Room temperature----->40 ℃    about 45minutes(Nonlinear no-load, about 1.2℃/min)
Indication Resolution(display accuracy)
Temperature: ±0.01℃
Control Precision(test precision)
Temperature: ±0.5℃
Distribution Accuracy (uniformity)
 
Temperature: ±1.0℃
Environment Temperature Range
5℃-28℃
Relative Humidity
≤85%RH
Supply Voltage
220V AC(±10%),protective grounding wire, ground resistance ≤4Ω
Frequency 
50±0.5Hz
Power Switch
Air-break switch (40A,  knife switch or outlet is forbidden to use)
Power 
8KW
Humidification Water
Resistivity ≥500MΩ
Requirement of Ambient Air Quality
Do not contain high concentration dust and flammable gas or dust,and no strong electromagnetic radiation source nearby.
Power Supply
AC1¢3W 380V 50/60HZ, 
Allowed value of voltage change: ±10% of rated voltage
Max Load Capacity
8KVA
Maximum Operating Current
30A