Key Specification
|
Model
|
OTS-A05
|
|
Dimensions of the Working Chamber
|
1000*1000*800 mm (W×H×D)
|
|
External Dimension of Chamber
|
1500*1860*1370MM (W×H×D)
|
|
Temperature Range
|
-50-200℃, can be set
|
|
Rate of Temperature Increase
|
Room temperature----->150℃ about 40minutes(Nonlinear no-load, about 3.5℃/min)
|
|
Room temperature----->40 ℃ about 45minutes(Nonlinear no-load, about 1.2℃/min)
|
|
|
Indication Resolution(display accuracy)
|
Temperature: ±0.01℃
|
|
Control Precision(test precision)
|
Temperature: ±0.5℃
|
|
Distribution Accuracy (uniformity)
|
Temperature: ±1.0℃
|
|
Environment Temperature Range
|
5℃-28℃
|
|
Relative Humidity
|
≤85%RH
|
|
Supply Voltage
|
220V AC(±10%),protective grounding wire, ground resistance ≤4Ω
|
|
Frequency
|
50±0.5Hz
|
|
Power Switch
|
Air-break switch (40A, knife switch or outlet is forbidden to use)
|
|
Power
|
8KW
|
|
Humidification Water
|
Resistivity ≥500MΩ
|
|
Requirement of Ambient Air Quality
|
Do not contain high concentration dust and flammable gas or dust,and no strong electromagnetic radiation source nearby.
|
|
Power Supply
|
AC1¢3W 380V 50/60HZ,
Allowed value of voltage change: ±10% of rated voltage
|
|
Max Load Capacity
|
8KVA
|
|
Maximum Operating Current
|
30A
|