Product Main

Specifications

GD-3536-1 Cleveland Open Cup Flash Point Tester 
Brief Introduction:
  • This instrument is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester
  • It is applicable to all petroleum products with flash points above 79°C and below 400°C except fuel oils.
Main technical features
1. It adopts technology of single chip machine and LCD screen. English man-machine interface.The heating device is continuously stepless adjustable.Easy to set parameters.
2. The LCD screen has prompt menu, prompt type input for operation interface. It shows set parameters and real-time display sample temperature and other parameters. Press the record key when flash point appearing. The screen will display and save flash point value.
3. It is newly design,having small structure. Stainless steel working table make it elegant and beautiful. The operator can do determination as long as connecting with coal gas or other civil gas. Easy to use and results are exact.
4. It is equipped with wind-shelter and flame extinguishing cover which are accord with requirements of test.
Accurate heating rate.The instrument can do test automatically.
5. Operator only need to observe the flash point appearing. The cost performance is high.
Main technical specifications and parameters
1. Power supply:  AC(220±10%)V,50Hz.
2. Heating device: Electric furnace heating, no naked fire, explosion prevented.The power is adjustable from 0W to 600W. The max heating temperature can reach 400°C.
3. Temperature control: Single chip machine.The heating rate can meet requirements of standards GB/T 3536-2008 and ASTM D92.
4. Temperature display: LCD screen shows temperature parameters. The range is 0°C~400°C, accuracy is 0.1°C.
5. Flash point detecting device: scanning the flash point automatically.
6. Temperature sensor: RTD, PT100
7. Igniting device: (1) Ignition source: coal gas (or civil gas,the same below)
                          (2) Flame diameter is 3.2mm~4.8mm
8. Ambient temperature: (-10∼50)°C
9. Relative humidity: ≤85%
10.Overall power consumption: ≤ 650W
11.Overall dimension: 340mm*320mm*450mm (Temperature sensor is included) 
Same series products:
Item GD-267 GD-3536 GD-3536B GD-3536A GD-3536D
Testing Range ambient~360°C   ambient~400°C   ambient~350°C   ambient~300°C   ambient~400°C
Repeatability ≤4 °C ≤8 °C    ≤4 °C   ≤8 °C ≤8 °C
Fire point  Can Test  Can  No  No  Can 
Automatic up/down  No  No  No Yes  Yes 
Automtic detect  No  No  Yes  Yes  Yes 
With micropriter  No  No  Yes  Yes  Yes 
Communication port  No  No  No  No  Yes 
Reproductility  ≤6 °C   ≤16 °C   ≤6 °C   ≤16 °C   ≤16 °C  
Screen Display  No  No  No  Yes  Yes