Specifications
GD-261D Fully-automated Pensky-Martens Flash Point Test is made as per ASTM D93 to measure the flash point of volatile oil.
Summary
This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products in the temperature range from 40°C to 360°C.
Main technical specifications and parameters
1. Power supply:AC(220±10%)V, 50Hz
2. Temperature measurement:Range: room temperature~300°C
Repeatability: ≤3°C
Reproducibility:≤6°C
Accuracy:0.1°C
3. Heating rate:Procedure A:(5~6)°C/min
Procedure B:(1~1.6)°C/min
Automatic control and manually adjustable.
4. Stirring rate:(1) Procedure A: (90~120)RPM
(2) Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5. Igniting mode:Electric ignition. Gas flame diameter:3.2mm~4.8mm
6. Working condition:ambient temperature:(10~40)°C
relative humidity:≤80%
7. Total power consumption:≤500W
8. Overall dimension: 520mm×360mm×310mm
Pakage
Item
|
Name
|
Unit
|
Qty
|
1
|
GD-261D Fully-automated Pensky-Martens Flash Point Test
|
Set
|
1
|
2
|
Oil cup
|
Cup
|
1
|
3
|
Power line(250V 6A)
|
Piece
|
1
|
4
|
Printing paper
|
Roll
|
1
|
5
|
Igniting gun
|
Piece
|
1
|