GD-2122B Automatic Water Content Karl Fischer Titrator
Brief Introdcution:
- The instrument is designed and made as per the National Standard of People's Republic of China GB7600 Test Method for Water Content of Transformer Oil in Service (Coulometric Method), and the Industry Standard of People's Republic of China SH/T0246 Test Method for Water Content of Light Petroleum Products (Electrometric Method).
- It is suitable to determine moisture in the liquid petroleum products.
- It adopts the microcomputer controlling technology, the instrument has the characteristics of rapid analysis, high precision, LCD indication, and automatic printing.
- It also has the function of self-diagnosis and menu selection.
- it is a multifunctional, convenient, automatic measurement and analysis instrument.
Main technical specifications and parameters:
1. Titration mode: coulometric titration controlled by a microcomputer
2. Measurement range: 1 μg~100 mg (Representative value is 10 μg~10 mg)
3. Electrolysis control: It can control the electrolysis current automatically (the maximum current is 400 mA)
4. Titration speed: 2 mg per minute (maximum)
5. Accuracy:
(1)10 μg~1 mg: ±3 μg (not including sample filling error)
(2) It will be 0.3% for more than 1 mg (not including sample filling error)
6. End point: data will be shown on the instrument, buzzer will alarm, and the indication light of end point will be on
7. Date: Year Month Day Hour Minute Second
8. Printer: 16 characters pin printer, width of paper 44 mm
9. Power supply: AC 220V±10%, 50 Hz±5%
10. Power: 60 W
11. Ambient temperature: 5°C~40 °C
12. Relative humidity: ≤85%
13. Size: 290mm×380mm×120mm
14. Weight: 9 kg
1. Titration mode: coulometric titration controlled by a microcomputer
2. Measurement range: 1 μg~100 mg (Representative value is 10 μg~10 mg)
3. Electrolysis control: It can control the electrolysis current automatically (the maximum current is 400 mA)
4. Titration speed: 2 mg per minute (maximum)
5. Accuracy:
(1)10 μg~1 mg: ±3 μg (not including sample filling error)
(2) It will be 0.3% for more than 1 mg (not including sample filling error)
6. End point: data will be shown on the instrument, buzzer will alarm, and the indication light of end point will be on
7. Date: Year Month Day Hour Minute Second
8. Printer: 16 characters pin printer, width of paper 44 mm
9. Power supply: AC 220V±10%, 50 Hz±5%
10. Power: 60 W
11. Ambient temperature: 5°C~40 °C
12. Relative humidity: ≤85%
13. Size: 290mm×380mm×120mm
14. Weight: 9 kg
| All The Oils Our Equipments Can Test | |
| No. | Test Sample |
| 1 | Petrol |
| 2 | Jet Fuel |
| 3 | Kerosene,Coal oil |
| 4 | Diesel oil |
| 5 | Gas turbine liquid fuel |
| 6 | Solvent oil |
| 7 | Lubricating oil |
| 8 | Lubricating grease,Grease |
| 9 | Transformer oil,Insulating oil |
| 10 | Hydraulic oil |
| 11 | Fuel oil(heavy oil) |
| 12 | Paraffin wax |
| 13 | Oils for Electric Power |
| 14 | New energy(biofuel) |
| 15 | Liquefied petroleum gas (LPG) |
| 16 | Asphalt,Bitumen |
| 17 | Laboratory instrument |
| 18 | Farm chemical |
| 19 | Building bitumen |
| 20 | Petroleum asphalt |
| 21 | Engine cooling liquid |
| 22 | Base oil,Lube oil |
| 23 | Naphthalene products |
| 24 | Industrial Aromatic hydrocarbonand related materials |