Product Main

Specifications

Cryostat Microtome RD-2230 ,Chamber temperature can drop within 120 minutes  of turning  
the unit on. The stand-by mode can be controlled between -4ºC and -9ºC automatically.  
The system returns to working temperature within 10 minutes of initiation from stand-by mode. 
 
 
Features: 
 
v Large stainless steel freezing chamber and scrap collector 
 
v Up to of 10 specimen pieces 
 
v LCD screen shows thickness, function, day, time, temperature, and time on/off 
 
v UVC sterilize function 
 
v Fast Freezing 
 
v Two defrost methods 
 
v Control panel with locking function 
 
v Locking handwheel 
 
v Sensor self-diagnosis function 
 
 
 
Technical Specification 
 
 
 
v Section thickness : 1-80μm  
 
                1μm increment from 1 to 20μm  
                         2μm increments from 20 to 40μm  
                        5μm increments from 40 to 80μm  
 
v Trimming thickness :  10-400μm  
 
                       5μm increments from 10 to 50μm  
 
                     10μm increments from 50 to 100μm 
 
                      50μm increments from 100 to 400μm  
 
v Chamber temperature ;   -10°C~-30°C  
 
v  Freezing shelf temperature : -45°C  
 
v  Peltier temperature   -55°C  
 
v  Number of freezing station 10  
 
v Peltier number    2  
 
v Refrigerant   R404A 300g±10g  
 
v Retraction 20µm (on or off) 
 
v Horizontal stroke  20mm  
 
v Vertical stroke  60mm  
 
v Maximum specimen size  Φ35mm  
 
v Voltage  220V±10%(can be customized)  
 
v  Frequency 50Hz  
 
v Power  850W  
 
v Weight  130KG