Product Main

Specifications

Main body

High-strength engineering plastic rack, light,

beautiful and corrosion resistant
Scope of application
adult, child
Gas source

O2: 0.27--0.55MPa    N2O: 0.27--0.55MPa

Flow meter
O2 : 0.1--1.0L/min   1.1--10L/min
N20: 0.1--1.0L/min   1.1--10L/min
Air : 0.1--10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply
25--75L/min
Low oxygen pressure alarm

There will be sound alarm when the oxygen pressure < 0.2MPa

Vaporizer

it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0--5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Respiratory circuit

working mode: all-close, semi-close, semi-open

APL: 0.5--7 kPa
Respiratory Bellows

bellows for adults, tidal volume range:0--1500 ml

bellows for children, tidal volume range:0--300 ml

anyone can be chosen for application as required by the customer

Ventilator
Display mode

High-definition 8.4” TFT color LCD screen display

Ventilation mode
IPPV, SIPPV, SIMV, PEEP, MANUAL,
SIGH(IMV, VCV, PCV)
Ventilation parameters
Tidal volume  50--1500 ml (20--1500 ml)
Rate         1--99 bpm
SIMV rate     2--20 bpm
I:E           2:1--1:8(4:1--1:8)
Inspiratory trigger pressure   -1.0--2.0 kPa
PEEP         0--2.0 kPa
Pressure range 1.0--6.0 kPa
SIGH        1.5 times the inspiratory time
(70-150 adjustable)
Inspiratory platform  0--50%
Parameters for ventilation monitoring

tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform,PEEP,inspiratory trigger pressure, Inspiratory platform

(autonomous inspiratory rate, lung compliance, airway resistance,flow-volume loop, pressure-volume loop)

Monitoring of oxygen concentration

21%--100%
Security Alarm System
Oxygen Concentration alarm
upper limit setting range  21%--100%
low limit setting range    10%--80%
Airway pressure alarm
upper limit setting range  0.3--6.0 kPa
low limit setting range     0.2--5.0 kPa

Per-minute ventilation volume alarm

upper limit setting range   3.0--30L/min
low limit setting range   1.0--10L/min
Sustained high-pressure alarm

it will give alarm when stress have consistently been higher than 2.5 kPa

Suffocation alarm

it will give sound and light alarm if there is no tidal volume input for 15 seconds

Power alarm
Power
AC 220V±10%     50Hz±1Hz
(UPS, Storage battery)
(Mechanical Arms)

(End-expiratory CO2 monitoring)

*Items within brackets are optional.