Main body
|
High-strength engineering plastic rack, light,
beautiful and corrosion resistant
|
|
Scope of application
|
adult, child
|
|
Gas source
|
O2: 0.27--0.55MPa N2O: 0.27--0.55MPa |
|
Flow meter
|
O2 : 0.1--1.0L/min 1.1--10L/min
N20: 0.1--1.0L/min 1.1--10L/min
Air : 0.1--10L/min
|
|
O2, N2O linkage and N2O Stopper
|
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
|
|
Flow rate of rapid oxygen supply
|
25--75L/min
|
|
Low oxygen pressure alarm
|
There will be sound alarm when the oxygen pressure < 0.2MPa |
|
Vaporizer
|
it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0--5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
|
Respiratory circuit
|
working mode: all-close, semi-close, semi-open
APL: 0.5--7 kPa
|
|
Respiratory Bellows
|
bellows for adults, tidal volume range:0--1500 ml bellows for children, tidal volume range:0--300 ml anyone can be chosen for application as required by the customer |
|
Ventilator
|
|
|
Display mode
|
High-definition 8.4” TFT color LCD screen display |
|
Ventilation mode
|
IPPV, SIPPV, SIMV, PEEP, MANUAL,
SIGH(IMV, VCV, PCV)
|
|
Ventilation parameters
|
Tidal volume 50--1500 ml (20--1500 ml)
Rate 1--99 bpm
SIMV rate 2--20 bpm
I:E 2:1--1:8(4:1--1:8)
Inspiratory trigger pressure -1.0--2.0 kPa
PEEP 0--2.0 kPa
Pressure range 1.0--6.0 kPa
SIGH 1.5 times the inspiratory time
(70-150 adjustable)
Inspiratory platform 0--50%
|
|
Parameters for ventilation monitoring
|
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform,PEEP,inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow-volume loop, pressure-volume loop) |
|
Monitoring of oxygen concentration |
21%--100%
|
|
Security Alarm System
|
|
|
Oxygen Concentration alarm
|
upper limit setting range 21%--100%
low limit setting range 10%--80%
|
|
Airway pressure alarm
|
upper limit setting range 0.3--6.0 kPa
low limit setting range 0.2--5.0 kPa
|
|
Per-minute ventilation volume alarm |
upper limit setting range 3.0--30L/min
low limit setting range 1.0--10L/min
|
|
Sustained high-pressure alarm
|
it will give alarm when stress have consistently been higher than 2.5 kPa |
|
Suffocation alarm
|
it will give sound and light alarm if there is no tidal volume input for 15 seconds |
|
Power alarm
|
|
|
Power
|
AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
|
|
(Mechanical Arms)
|
|
|
(End-expiratory CO2 monitoring) |
|
|
*Items within brackets are optional. |
|