Product Main

Specifications

Adopting low differential sensor technology, the third-generation differential pressure transmitter are insensitive to media and without pressure drift
Applicable Medium  Air or moncombustible gas filtering usually can be ignore
 Accuracy  1% ~ 2% (optional )
 Measure range  Min 0~25 Pa  Max 0~3K Pa  
 Temperature range: -10°C ~ 50°C
 Temperature drift   0.05%
 Pressure limit  1PSI max. operation .10PSI burst
 Power supply  10~35 VDC (2-wire) ;
 17~36 VDC or isolated 21.6~33 VAC (3-wire)
 Output  4-20mA ( 2-wire ) ; 0-10V ( 3-wire )
 Response time  Field adjustable 0.5~15 Sec ; 0~5V ( 3-wire )
 Zero point and meas-
 ure range adjustable
 Digital input keys
 Loop impedance  Current Output 0~1250Ω Max; Voltage output 1000Ω Min load
 Power consumption  40mA max
 Display(optional)  LCD display with 4 letters
 Electrical connection  4~20mA, 2-wire ;  Connecting : 1/2” NPS connector
 Accessory  Sealed tube for 4~7.5mm diameter cable
 Process Connection  3/16” ( 5mm ) ID tube , Max OD 9mm
 Protection Class  Ip65
 Installation  Vertical installation