Product Main

Quick Details

7" LCD display: Capture dynamic waveforms

Specifications

Specifications
small sie 
7" LCD display 
ECG, NIBP, SpO2, RESP 
Capture dynamic waveforms 
IBP and Etco2 is optional

 

Specification of Patient Monitor

Features: (ZD-120)

7" color TFT display

Unique human voice alarm

Maximum 720 hour graphic and tabular
trends

Maximum 12 hours of all waveforms

Support USB data output

System software can be upgraded through
network

Wire and wireless network

Standard
configuration: 

ECG, NIBP, SPO2, TEMP, RESP, PR 
Optional: 

Dual TEMP 

· Dual IBP 

· EtCO2 (side stream) 

· Nellcor SpO2 module 

· Thermal Printer

Specification:


Item No.


TB-800A


Screen Size:


7"


ECG

 


lead mode;I,II,III,AVR,AVL,AVF,V
Gain:X0.25,X0.5,X1,X2
Heart Rate:15-300BPM(Adult) 15-350BPM(Neonatal) 
Resolution:1 BPM
Accuracy:+-1%
ST measurement Range:-2.0-+2.0mV
Accuracy:+-10%
Pacemaker:Yes
Sweep Speed:12.5 mm/s,25mm/s,50mm/s
Bandwidth:Diagnostic:0.05--130Hz
Monitor: 0.5--40Hz
Surgery: 1-20Hz


Spo2

 

Measure Range:0-100%
Resolution:1%
Accuracy:2%(70-100%)
Pulse Rate:20-250BPM
Resolution:1BPM
Accuracy:+-3BPM


NIBP

 

Work
mode:Manual,Automatic,Continuous
Unit:mmHg,Kpa
Measure Range:
Adult:
SYS: 40-280mmHg
DIA: 10-225mmHg
MEAN:20-240mmHg
Pediatric:

SYS:40-220mmHg
DIA: 10-160mmHg
MEAN:20-170mmHg
Neonatal:

SYS:40-135mmHg
DIA:10-100mmHg
MEAN:20-110mmHg

Resolution:1mmHg
Accuracy:+-5mmHg


RESP

 

Measure Range:0-150BrPM
Resolution:1BrPM
Accuracy:+-2%


TEMP

 

Measure Range:0-50C
Resolution:0.1C
Accuracy:+-0.1C