500mA Surgical Radiography & Fluoroscopy X-ray Machine (RF-5000A)

Technical Specification
Item Contents
Power Supply Voltage: 380V± 38V Frequency: 50Hz± 1Hz Capacity: ≥ 40kVA, Internal Resistance: ≤ 0.17Ω

Photography
X-ray Tube Voltage: 44kV~125kV, totally 41 grades
X-ray Tube Current: 32mA~500mA adjustable with different grades
Exposure time: 0.02s-4s, totally 21 grades; Control interface: Touch-screen
Digital X-ray High Voltage System

Fluoroscopy
X-ray Tube Voltage: 44kV~110kV continuous adjustable with 2kV each grade
X-ray Tube Current: 0.5mA~5mA continuous adjustable
Digital controlled X-ray tube
Type: XD51-20, 40/125; Power output: 40kW(max) 20kW(min)
Anode thermal capacity: 100kJ; Assembly thermal capacity: 900kJ (1250kHU)
Rotary anode speed: 2700rpm; Tube focus: 2.0 (max) 1.0 (min)
Micro-controlled digital upper tube remote-controlled surgical bed
Movement range: 90° ~0° ~25°
Bed transverse movement range distance: ± 110mm
Portrait movement of photography frame and track spot film: ≥ 720mm
Photography dimension of whole film: 8"× 10"~14" × 17"
Spot film division: Whole film, 2-division film, 3-division film, 4-division film
Method to control bed: Remote control, bed-control, Soft boot and stay of frequency conversion
Beam applicator: Electrodynamics multi-pin
Bed-applied fixed bucky grid: Grid density: 103L/INCH
Grid ratio: 10: 1, Focus distance: 120cm, 15" × 18"

Digital TV system
EGA: 9" Toshiba, High resolution CCD (digital camera); Imported low-light 0.47 MEGA PILEXS
Monitor: 14" medical high definition monitor, horizontal central resolution 1000 thread, edge 800
Thread, video wideband 12.5MHz, 50bit, 25 pictures per second
Picture-process system: High definition and non-interlaced output, 8-grade noise reduction, 7 pictures in storage, LIH, mirror image rotation in vertical and horizontal, positive-negative image; OSD (monitor display)