Product Main

Quick Details

Place of Origin: China (Mainland)
Brand Name: Gold
Model Number: GD-3536D
Usage: Cleveland Open Cup Flash Point Tester
Testee: Petroleum Products
Method: Cleveland Open Cup Methods
Standard: GB/T3536, ISO2592, ASTMD92
Size: 48cm*40cm*30cm
Weight: 8.5kg
overseas warranty: 12 month
Packing: wooden case
Delivery time: 5-7 days

Specifications

GD-3536D.jpg 

 

Application:

The instrument is designed and made as per the National Standard GB/T3536 "/ISO2592/ASTM D92Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Methods), and T0651 "Asphalt Flash Point and Fire Point Test (Cleveland Open Cup Methods)" in the Industry Standard JTJ052 Specification and Test Methods of Bitumen and Bituminous Mixtures for Highway Engineering. It is used to determine flash point of petroleum products and bituminous materials, excepting fuel oil and materials having a flash point of lower than 79 centigrade.
Adopt microprocessor techniques and equipped with advanced system software, it can control temperature, apply flame, ignite, detect flash point, and print out test data automatically. It has a 6 digits display. 4 digits for temperature display and the other 2 digits for time display.

 

Main technical characteristics:

1. The instrument is an automatic flash point and fire point tester. It is the best choice among the similar instruments.

 

2. The instrument adopts SCM controlling technologies, colored LCD, man machine conversation interface, and blank buttons. It has prompt menu to help inputting expected flash point, sample serial number, barometric pressure, test date, and other parameters.

 

3. The instrument adopts desktop structure. It is advanced, friendly, convenient, and accurate. It has been widely used in railway, aviation, power plant, petrochemical industry, relative college and universities, scientific research institutes, and measurement and inspection department for petroleum product flash point and fire point test.
Main technical specification and parameters:

1. Power supply: AC 220V±10%, 50 Hz;

 

2. Temperature measurement:

(1) Measurement range: room temperature~400°C;

(2) Repeatability: ≤4°C;

(3) Reproducibility: ≤8°C;

(4) Resolution: 0.1 °C;

(5) Accuracy: 0.5%;

 

3. Temperature rising speed: as per GB/T3536;

 

4. Ignition mode: by electron igniter; Gas flame of 3~4 mm for flame application;

 

5. Ambient temperature: 10~40°C;

 

6. Relative humidity: ≤80%;

 

7. Total power consumption: not more than 400 W;

Full set:

 

No.

Name

Unit

Quantity

Remark

1

Main unit

Set

1

 

2

Cleveland oil cup

Piece

1

We can provide more if you required.

3

Ignition device

Set

1

 

4

Thermometer

Piece

1

-6°C~400 °C, graduated scale is 2 °C

5

Thermometer support

Set

1

 

6

Power supply wire (250 V 6 A)

Piece

1

 

7

Fuse 5 A (Φ5×20)

Piece

2

 

8

Latex tube Φ6

Meter

1.5

Could not be used as coal gas tube

9

Joint for coal gas tube

Piece

1

Connect coal gas tube and latex tube

10

Extinguishing cover 

Piece

1

 

 

The same series products: 

Model

GD-267

GD-3536

GD-3536-1

GD-3536A

GD-3536D

Range

Ambient~360°C

Ambient~400°C

Ambient~400°C

Ambient~300°C

Ambient~400°C

Repeatability

≤4°C

≤8°C

≤4°C

≤8°C

≤8°C

Reproducibility

≤6°C

≤16°C

≤6°C

≤16°C

≤16°C

Fire point

Y

Y

N

N

Y

Automatic up/down

N

N

N

Y

Y

Automatic detect

N

N

Y

Y

Y

With microprinter

N

N

Y

Y

Y

Communication port

N

N

N

N

Y

Display

N

N

N

Y

Y

 

 

 2.jpg3.jpgISO.jpg