Product Main

Specifications

3 Parts Auto Hematology Analyzer

Usage
It is used to detect the parameters of the series of red blood cells and hemoglobin, platelet and leukocyte in the blood. It is an in-vitro diagnostic equipment which is applied to quantitative analysis of blood cells of clinical and laboratory.

Feature

1. The principle of this machine is electrical impedance (EI).

2. The machine is mainly used for the quantitative measurement of the blood in the 19 parameters and 3 histograms.

3. This machine use Screw transmission technology, it ensures the testing result more accurate.

4. It can save 20, 000 test results and the data can be searched by date, ID, SN.

5. The size of LCD display is 8".

6. It has a built-in thermal printer.

7. It has four interfaces: RS232 joint, USB, PS2, Parallel port.

8. It can set up and edit the parameters of male, female, children and newborn.

Our other medical products including X-ray machine & accessories, Ultrasound scanner, Ventilator & Anesthesia machine & Patient monitor & Defibrillator, ECG machine, Hospital furniture, Fast diagnostic kits.

 

 



Test Items WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW—CV, RDW—SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram
Background Count WBC: ≤0.3×10 9 /L   RBC: ≤0.03×10 12 /L   HGB: ≤1g/L   PLT: ≤10×10 9 /L
Linearity Test Item Test Range Allow Difference of Linearity
WBC 0~6.0×10 9 /L ±0.3×10 9 /L
6.0~99.9×10 9 /L ±5%
RBC 0~0.99×10 12 /L ±0.05×10 12 / L
1.0~9.99×10 12 /L ±5%
HGB 0~99 g/L ±2g/L
100~300 g/L ±2%
PLT 0~99×10 9 /L ±10×10 9 /L
100~999×10 9 /L ±10%
Repeatability Test Item CV(%) Teat Range
WBC ≤2.0 7.0-15.0×10 9 /L
RBC ≤0.5 3.5-6.0×10 12 /L
HGB ≤1.5 110-180g/L
PLT ≤4.0 150-500×10 9 /L
MCV ≤0.5 80-110fL
Sample Volume Twig blood: 20μl
Dilution Ratio WBC/HGB: 1:500   RBC/PLT: 1:50000
Interfaces RS232 joint, USB, PS2, Parallel port
Test Speed 60 samples/hours
Continuous Work Time >8h