Trolley EMG/EP System (Evoked Potential function)
Model No.: NTS-2000-A
More Models of Trolley EMG/EP Systems for your optional:
A: Trolley EMG System
1) NTS-2000-A12: 2 channels trolley EMG System;
2) NTS-2000-A14: 4 channels trolley EMG System.
B: Trolley EMG/EP System (Evoked Potential function)
1) NTS-2000-A32: 2 channels trolley EMG/EP System (Evoked Potential function);
2) NTS-2000-A34: 4 channels trolley EMG/EP System (Evoked Potential function).
Features & Advantages
1)Examination protocols: EMG, NCV, SEP; (AEP, VEP, P300 is optional function for Evoked Potential);
2)Pre-amplifier: 2/4 channels optional, and it can be updated to 8 or 16 channels;
3)Fiber optical transmission and preeminent anti-interference capability;
4)Easy and fast operation with operation panel and footswitch;
5)Modular design: To make transportation and maintenance more easier, and reserve Magnetic stimulation interface.
Specification of Amplifier:
1)Sensitivity: 0.05UV/div~20000UV/div multistage control
2)Amplitude-frequency characteristic: 0.2Hz~10kHz
3)Noise: ≤ 0.4UV(RMS)
4)CMRR: ≥ 110dB
5)Input impedance: ≥ 1000MΩ
6)Scanning speed measure error: Between 1ms/div-5000ms/div