Product Main

Quick Details

Place of Origin: China (Mainland)
Brand Name: Gold
Model Number: GD-3536A
Usage: Cleveland Open Cup Flash Point Tester
Application: To Test The Flash point of Petroleum Products
Method: Cleveland Open Cup Methods
Standard: GB/T3536, ISO2592, ASTMD92
Size: 48cm*40cm*30cm
Weight: 8.5kg

Specifications

Application

The instrument is designed and made as per the National Standard GB/T3536 " Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Methods), and T0651 " Asphalt Flash Point and Fire Point Test (Cleveland Open Cup Methods)"  in the Industry Standard JTJ052 Specification and Test Methods of Bitumen and Bituminous Mixtures for Highway Engineering. It is used to determine flash point of petroleum products and bituminous materials, excepting fuel oil and materials having a flash point of lower than 79 °C.

The instrument is used to determine flash point of petroleum products. It has function of automatic program temperature controlling, LCD display, automatic barometric pressure calibration, and automatic modified value calibration. It can apply flame, ignite, detect and print out test data automatically. It is convenient and rapid. It is the best substitute product for imported one

Technical parameters

1. Power supply: AC 220 V±10%, 50 Hz;

2. Temperature measurement:

(1) Full scale: Ambient temperature to 300 °C;

(2) Repeatability: ≤8°C;

(3) Reproducibility: ≤16°C;

(4) Resolution: 0.1°C;

(5) Precision: 0.5%;

3. Basic parameters:

(1) Temperature rising speed: as per GB/T3536;

(2) Ignition mode: by an electric flame (gas flame is 3 to 4 mm);

4. Ambient temperature: 10~40°C;

5. Relative humidity: ≤80%;

6. Total power consumption: not more than 500 W;

 

 

Full set:

 

No.

Name

Unit

Quantity

Remark

1

Main unit

Set

1

 

2

Cleveland oil cup

Piece

1

We can provide more if you required.

3

Ignition device

Set

1

 

4

Thermometer

Piece

1

-6°C~400 °C, graduated scale is 2 °C

5

Thermometer support

Set

1

 

6

Power supply wire (250 V 6 A)

Piece

1

 

7

Fuse 5 A (Φ5×20)

Piece

2

 

8

Latex tube Φ6

Meter

1.5

Could not be used as coal gas tube

9

Joint for coal gas tube

Piece

1

Connect coal gas tube and latex tube

10

Extinguishing cover 

Piece

1

 

 

 

 

The same series products: 

Model

GD-267

GD-3536

GD-3536-1

GD-3536A

GD-3536D

Range

Ambient~360°C

Ambient~400°C

Ambient~400°C

Ambient~300°C

Ambient~400°C

Repeatability

≤4°C

≤8°C

≤4°C

≤8°C

≤8°C

Reproducibility

≤6°C

≤16°C

≤6°C

≤16°C

≤16°C

Fire point

Y

Y

N

N

Y

Automatic up/down

N

N

N

Y

Y

Automatic detect

N

N

Y

Y

Y

With microprinter

N

N

Y

Y

Y

Communication port

N

N

N

N

Y

Display

N

N

N

Y

Y