Product Main

Quick Details

Place of Origin: China (Mainland)
Brand Name: Refine Med
Model Number:

Specifications

Medical Device Infusion Pump JSB-1200

Features


1. Can be used for all standards of IV sets.
2. Automatic Calibration for IV sets.
3. Three working modes: Rate control mode, Time control mode and volumetric mode.
4. Built in rechargeable battery (Lithium-ion polymer battery).
5. Spec of the fuse: F1AL /250V, two (inside).
6. 8 hours internal Lithium-ion polymer battery backup.
7. Alarm functions: occlusion, empty bottle, low battery, rate error, end of infusion, door open, air in line etc.
8. Three levels of the occlusion limited pressure: high, middle and low.
9. Three levels of alarm sound: high, middle and low.
10. The infusion pump goes into KVO mode automatically as infusion is completed.
11. Purge and Bolus functions
12. Waterproof class: IPX4


Technical Specification


1. Flow range: 1.0~99.9ml/h, 0.1ml/h increment, 100~1200ml/h, 1ml/h increment;
or 1~400drops/min, 1 drop/min increment.
2. Flow precision: ± 5%
3. Target Volume: 0~9999ml, 1ml increment
4. Time Limit Range: 1~1999min, 1min increment
5. Purge Rate: 600 ml/h.
6. Bolus : Adjustable from 10~1200ml/h, in steps of 0.1ml/h (flow rate < 100ml/h) and 1ml/h(flow rate > 100ml/h)
7. KVO rate : if the flow rate ≤10 ml/h, KVO is 1ml/h. Otherwise KVO is 3ml/h.
8. Power supply: AC100~240V, 50/60Hz, Internal Lithium-ion polymer battery: DC 11.1V, ≥1600mAh. The battery backup is approx. 8 hours (when using a, fully charged battery and running at 20 drops/min with any standard IV Set).
9. Power consumption: ≤25VA
10. Class: I
11. Type: CF
12. Size: 125mm × 125mm × 230mm
13. Net weight: approx. 2.2kg
14. Operating condition: Ambient temperature: 5~40 degree C,
Relative humidity: 20~90%
15. Storage condition: Ambient temperature: -10~55degree C
Relative humidity : 10~90%.