Product Main

Quick Details

Place of Origin: China (Mainland)
Brand Name: GOLD
Model Number: GD-264A

Specifications

Purpose:

The instrument is suitable to determine acid number of transformer oil and turbine oil.It is designed and made as per the National Standard GB/T264-83 " Test Methods for Acid Number of Petroleum Products" . It takes a blue LCD for display. It can make determination, print test data automatically using a microprocessor. Test procedures are automatically controlled by the microprocessor. Operation procedures are intuitive and simple. Adopt a magnetic stirrer, it is safe, reliable, delicate, and there is less noise during test procedure.

 

Technical parameters:

1. Power supply: 220V±10, 50Hz;

2. Power consumption: 100W;

3. Measurement range: 0.0010.999 mg KOH/g;

4. Resolution: 0.001 mg KOH/g;

5. Test accuracy:

(1) Acid number 0.1: ±0.005 mg KOH/g;

(2) Acid number ≥0.1: 5% of displayed value;

6. Repeatability: 0.004 mg KOH/g;

7. Ambient temperature: 10℃40℃;

8. Relative humidity: 75;