Specification
ECG
Full lead: I, II, III, AVL, AVR, AVF and V
Gain: 5mm/mV, 10mm/mV, 15mm/mV, 20mm/mV and 25mm/mV
Scanning speed: Changeable with three grades of 12.5mm/s, 25mm/s and 50mm/s
Band width: 0.05Hz-100Hz(3dB)
Heart rate:
Scope: 15 times/minute to 300 times/minutes
Accuracy: ±1 time/minute
Variation of heart beat LORENZE drawing: Analysis on heart beat data for the current 5 minutes with histogram during RR period.
Arrhythmia: Twenty kinds of common abnormal ECG can be analyzed, with ECG data of 1000 times of arrhythmia kept and with 10 seconds kept for each time.
Accuracy: Error rate < 20%
Undetected rate < 20%
ST section
Scope: ±0.8mV
Precision : < ±0.05mV
SpO2
Detection scope: 0%-100%
Accuracy ±1% 90%-100%
±2% 70%-89%
NIBP
Theory Electric oscillator
Cuff gassing time less than 15 seconds
First-time gassing pressure: dynamic state 24kPa(180 mmHg)
Manual selection: 9.3kPa(70 mmHg), 13.3kPa(100 mmHg), 16.0kPa(120 mmHg)
18.6kPa(140 mmHg), 20.0kPa(150 mmHg), 21.3kPa(160 mmHg),
Automatic detection period: 1 minute to 240 minutes
Stop size: 1 min(1 min -10min), 5min(10 min -30min), 10min(30 m
in -90min) and 30min(90 min -240min)
Detection scope:
Systolic blood pressure: 4kPa -34kPa(30mmHg -255mmHg)
Diastolic pressure 2kPa -29.3kPa(15mmHg -220mmHg)
Mean pressure 2.7kPa~31.3kPa (20mmHg~235mmHg)
Detection accuracy
Static pressure: ±3mmHg
Pulse rate: ±2%
Resp (Optional)
Chest impedance method or concentration of carbon dioxide at end of respiration
Scope of chest impedance method 0 time/minute – 100 minutes/minute
Precision of chest impedance method time/minute
Detection scope of carbon oxygen at end of respiration 3mmHg-75mmHg; 0.4%vol-9.9%vol
Detection scope of respiration rate: 2 times/minute ->60 times/minute(including baby)
Detection precision of respiration rate: ±1 time/minute
Temp (Optional)
Single/double path
Type: Vascular/surface temperature
Scope: 20℃~50℃
Accuracy: ±0.2℃
Invasive blood pressure (optional)
Single/double path
Type: direct and continuous detection
Scope: -6.67kPa(-50mmHg)—46.67kPa(350mmHg)
Precision: below 13.33kPa(100mmHg) ±0.27kPa(2mmHg) 13.33k
Pa(100mmHg) above ±2%
Frequency response: DC-12Hz
Input resistance: not less than 500k?
Sensitivity of sensor: 5μV/V/mmHg
Motivation voltage: 5V (DC)
Accuracy of sensor: ±2%Isolation voltage: 4 kV
|