This Flash Point Test Machine can determine the flash point and fire point of lubricating oil and Dark petroleum products. 

 
Product Description

   GD-267 Manual Flash Point Test Machine by Open Cup

 

Purpose

GD-267 Open Cup Flash Point Test Machine is designed and made as per the National Standard GB/T267 " Test Methods for Flash Point and Fire Point of Petroleum Products (Open Cup Methods)"  and Industry Standard of Petrochemical SH/T0318 " Technical Requirements for Open Cup Flash Point Tester" .
GD-267 Open Cup Flash Point Test Machine is used to determine the flash point and fire point of lubricating oil and Dark petroleum products.
 
Technical parameters
1. Power supply: AC 220 V±10%,50Hz;
2. Outer crucible:
(1) Material: Carbon structural steel;
(2) Diameter of upper opening: Φ100±5 mm;
(3) Height: 50±5 mm;
(4) Diameter of bottom: Φ56±2 mm;
3. Inner crucible:
(1) Material: Carbon structural steel;
(2) Diameter of upper opening: Φ64±1 mm;
(3) Height: 47±1 mm;
(4) Diameter of bottom: Φ38±1 mm;
(5) Scale line: there is a line 12 mm away from the rim of upper opening and a line 18 mm away from the rim of upper opening.
4. Gas tube: the diameter of nozzle isΦ0.8~Φ1 mm. The inner surface of tube is smooth and it can adjust flame to 3 to 4 mm.
5. Heater: Heated by an electric furnace. Heating power is 1000 W and adjustable.
6. Thermometer: 0~360 °C; each scale division is 1 °C. It conforms to GB/T514.
7. Ambient temperature: ≤35 °C;
8. Relative humidity: ≤85%;
9. Total power consumption: not more than 1100 W;
 