Product Main

Quick Details

Model Number: HPC300
Brand Name: STARK
Place of Origin: China (Mainland)
keyword: dust particle counter
title: HPC300 portable dust particle counter with USB
name: portable dust particle counter
Particle size: 0.3,0.5,5um(0.3-10um)
Power: Electronic,lithium-ion battery
Weight: dust particle counter< 800 grams (including battery)
packge: Standard export carton for dust particle counter
Output interface: dust particle counter with USB ,RS232
factory location: 2 hours from shanghai
certificate: dust particle counter with CE

Specifications

SDK  HPC300 portable dust particle counter with USB

Product introduction

SDK HPC300 portable dust particle counter with USB can be used in measuring particles suspended in the air in clean environment applications such as microelectronics, fine mechanics, optics, pharmaceutical,medical device, food processing, aerospace as well indoor/outdoor air quality. 

  

 The SDK-HPC300   dust particle counter was designed in the USA and is in compliance with the international standards (JIS B9925:1997 and ISO14644-1) and has CE certification. The HPC 300   dust particle counter is very unique compared to other manufacturers in the market; it features high sensitivity, multiple functional capabilities, and is slim,lightweight and very user friendly.

 Advantage

HPC300 portable dust particle counter with USB

 

1.Counting modes with cumulative differential /concentration /average / auto-repeat / timer

 

2.Up to 1500-data(450 sets)internal memory

 

3.Excess-count-limit warning USB interface for data downloading and upgrading

 

4.USB interface for data downloading and upgrading

Specification

  

Light Source

Laser diode (>100,000 hoursMTBF)

Sensitivity

0.3μm

Size Range

0.3μm~10μm

Channels

All channels are configurable by users (size selections from

0.3μm, 0.5μm, 0.7μm, 1.0μm,

2.0μm, 2.5μm, 5.0μm and 10μm

Counting

Efficiency

50±20% @0.3μm

100±10% (0.5μm)

Coincidence

Loss

<5% @70,000 Particles/Liter or

<5% @2,000,000 particles/ft3

Zero Count

<1 count per 5 minutes

Flow Rate

2.83 L /min (0.1cfm)

Sampling Time

1 sec ~ 59 min and 59sec (User defined)

Count Limit

Warning

FED STD 209E  (Class 1 ~ 100,000)

or ISO 14644-1 (Class 2 ~ 9)

Measurement

Single/repeat/Average

Error Indications

Count limit, loss of laser power, out of flow calibration (>±5%), insufficient battery

Interface

USB, RS232

Internal

Memory

1500 measurement data (450 sets)

Power

Li-ion polymer rechargeable battery

(7.4V/2800mAH) or 9VDC

AC Adapter (100~240V input)

Max. Operating

Time

Continuous operation > 5 hours with Li-ion battery

Dimensions

180 (H) ×93 (W) × 48 (D) mm

Weight

< 950 grams (including battery)

Environmental

Conditions

Operating: 5 ~ 45°C, < 90%RH Storage: -20 ~ 65°C, < 90%RH

Accessories