| Model Number: | PPC300 |
|---|---|
| Brand Name: | STARK |
| Place of Origin: | China (Mainland) |
| keyword: | particle size analyzer |
| title: | air particle size analyzer with USB |
| name: | particle size analyzer |
| packge: | Standard export carton |
| interface: | particle size analyzer with USB/RS232 |
| Weight: | particle size analyzer <3500grams |
| particle size: | 0.3um-5um |
| factory location: | 2 hours from shanghai |
| Power: | particle size analyzer with Electronic,lithium-ion battery |
| certificate: | particle size analyzer with CE |
Quick Details
Specifications
Product details
PPC300 portable air particle size analyzer with USB is used to measure the clean environment of dust particles per unit volume and distribution.
particle size analyzer takes advanced U.S. high-sensitivity miniature sensor technology independently developed set of aerodynamics, digital signal processing, optical and electrical integration of high-tech products.
The basic principle of the laser scattered by dust particles, the optical sensor outputs a pulse signal for digital signal processing, measure parameter settings.
The results, buttons, timing, printing, time, date, data storage are controlled and implemented by the built-in microcomputer (MCU).
And working status of the laser particle sensor.
The particle size analyzer is designed according to the international standard ISO and GMP It could detect 3 sizes (the user can arbitrarily set the test size).
Sampling time up to 59 minutes and 59 seconds, up to 1000 groups' sampled data can be stored in the built-in flash memory. And through USB or RS232 interface easy to download.
Features
?Simultaneously measures 3 user configurable particle sizes
?Counting modes with cumulative/ differential /concentration / average / auto- repeat / timer
?Up to 1000 sets internal memory
?Excess-count- limit warning
?USB/RS232 interface for data downloading and upgrading
?External digital temperature and humidity probe
?Ac/dc measurement print
?Flow meter included
Light Source
Laser diode (>100,000 hours MTBF)
Sensitivity
0.3μm
Size Range
0.3 ~ 25μm
Channels
User configurable on any channel with
0.1μm resolution (six default channel settings: 0.3, 0.5, 0.7, 1.0, 2.0, 5.0μm)
Counting
Efficiency
50±20% @ 0.3μm
100±10% @ 0.5μm
Coincidence
Loss
<5% @ 70,000 particles/liter
Zero Count
<1 count per 5 minutes
Flow Rate
2.83 L /min (0.1cfm)
Sampling Time
User defined: (up to 59m59s) and auto repeat (up to 99 times)
Count Limit
Warning
FED STD 209E (Class 1 ~ 100,000)
or ISO 14644-1 (Class 2 ~ 9)
Measurement
Single/repeat/Average
Error Indication
Count limit, loss of laser power, out of flow calibration (>±5%),
insufficient battery
Interface
USB, RS232, RJ45
Internal
Memory
6000 measurement data (1000 sets)
Power
Removable, rechargeable Lithium battery (7.4V/2800mAh) or 9VDC AC Adapter (100~240V input)
Max. Operating
Time
Continuous operation > 5 hours with
Lithium battery
Dimensions
Handheld: 185 (H) × 93 (W) ×46 (D)mm
Base unit: 152 (Dia.)× 97(H) mm
Weight
< 800 grams (including battery)
Environmental
Conditions
Operating: 5 ~ 45°C, < 90%RH Storage: -20 ~ 50°C, < 90%RH
Accessories
Main base unit, AC adaptor, iso- kinetic probe, USB data cable, remote sampling and data download software (CD), zero-count/purge filter, digital temperature and humidity sensor probe, traceable calibration certificate
Technical parameters