Product Main

Quick Details

Model Number: phs-98108
Place of Origin: China (Mainland)

Specifications

PHS-98108 tester family ATC Pen PH meter


SPECIFICATIONS

 

Technical data:

Measuring range: 0.00~14.00pH

Resolution: 0.1pH

Accuracy: ±0.1pH or 0.05pH

Calibration: manual, 2 points

ATC:   5-60 

Power: 4 x 1.5v alkaline (AG13)

Auto shut-off: Approx. 4minutes

Dimension: 175x41 x23mm

Weight: 90g

 

OPERATION

Remove the protective cap. Don't be worry if white crystals appear around the probe head. This is normal with pH electrodes and they dissolve when rinsed with water.

Turn the meter on by sliding switch on the top.

 Immerse into solution up to the max immersion level.

 Stir gently and wait until the display stabilizes.

 After use, switch off pH meter, use the water clean the electrode and replace the protective cap.

Large fluctuations in readings could be due to lack of calibration, dry electrode or rundown batteries.


  

CALIBRATION

Immerse the tester up to the maximum immersion level in pH 6.86 buffer solution

Allow the reading to stabilize and using a small screwdriver adjust the calibration trimmer to read 6.86

Calibration is now complete.

ALWAYS USE FRESH BUFFERS FOR CALIBRATION.


BATTERY REPLACEMENT

When the meter cannot be switched on or the display fades, pull out the battery compartment and replace all three 1.5V batteries, while paying attention to their polarity. Batteries should only be replaced in a safe area using the battery type specified in  instruction manual.(1.5 Alklaine or AG13)