Product Main

Specifications

Specifications
Voltage DC
Maximum Voltage:  1000V
Accuracy:  Fluke 83 V: ±(0.1%+1)
   Fluke 87 V: ±(0.05%+1)
Maximum Resolution:  Fluke 83 V: 100 μV
   Fluke 87 V: 10 μV
Voltage AC
Maximum Voltage:  1000V
Accuracy:  Fluke 83 V: ±(0.5%+2)
   Fluke 87 V: ±(0.7%+2) True RMS
AC Bandwidth  Fluke 83 V: 5kHz
   Fluke 87 V: 20kHz*
   *with low pass filter; 3db @ 1kHz
Maximum Resolution:  0.1 mV
Current DC
Maximum Amps:  10A (20 A for 30 seconds maximum)
Amps Accuracy:  Fluke 83 V: ±(0.4%+2)
   Fluke 87 V: ±(0.2%+2)
Maximum Resolution:  Fluke 83 V: 0.01 mA
   Fluke 87 V: 0.01 μA
Current AC
Maximum Amps:  10A (20 A for 30 seconds maximum)
Amps Accuracy:  Fluke 83 V: ±(1.2%+2)
   Fluke 87 V: ±(1.0%+2) True RMS
Maximum Resolution:  0.1 μA
Resistance
Maximum Resistance:  50 MΩ
Accuracy:  Fluke 83 V: ±(0.4%+1)
   Fluke 87 V: ±(0.2%+1)
Maximum Resolution:  0.1Ω
Capacitance
Maximum Capacitance  9,999 μF
Accuracy  ±(1%+2)
Maximum Resolution  0.01 nF
Frequency
Maximum Frequency  200 kHz
Accuracy  ±(0.005%+1)
Maximum Resolution  0.01 Hz
Duty Cycle
Maximum Duty Cycle  99.9%
Accuracy  ±(0.2% per khz +0.1%)
Maximum Resolution  0.1%
Temperature Measurement
Fluke 87 V, 87 V/E:  -200.0°C - 1090°C
   -328.0°F - 1994.0°F
   excluding probe
80 BK Temperature Probe
   -40.0°C - 260°C
   -40.0°F - 500°F, 2.2°C or 2% whichever is greater
Conductance
Maximum Conductance  60.00 nS
Accuracy  ±(1.0%+10)
Maximum Resolution  0.01 nS
Diode
Range:  3 V
Resolution:  1 mV
Accuracy:  ± (2 % + 1)
Duty Cycle Range
Accuracy:  Within ± (0.2% per kHz + 0.1 %)

 

Environmental Specifications
Operating Temperature
   -20°C to +55°C
Storage Temperature
   -40°C to +60°C
Humidity (Without Condensation)
   0% - 90% (0°C - 35°C)
   0% - 70% (35°C - 55°C)
Operating Altitude
   2000 m

 

Safety Specifications
Overvoltage category
   EN 61010-1 to 1000 V CAT III, 600V CAT IV
Agency Approvals
   UL, CSA, TüV, VDE listed

 

Mechanical & General Specifications
Size
   201 x 98 x 52 mm (with holster)
Weight
   355 g
Weigh With Holster and Flex-Stand
   624 g
Display
Digital:  6000 counts updates 4/sec.
   (Model 87 also has 19,999 counts in high-resolution mode.)
Analog:  32 segments, updates 40/sec.
Frequency:  19,999 counts, updates 3/sec at > 10 Hz.
Warranty
   Lifetime
Battery Life
Alkaline:  ~400 hours typical, without backlight
Shock
   1 Meter drop per IEC 61010-1:2001
Vibration
   Per MIL-PRF-28800 for a Class 2 instrument