Product Main

Quick Details

Model Number: TR600
Brand Name: Sharingtek
Place of Origin: China (Mainland)
Dynamic Range: 32/30dB, Single Mode, 1310/1550nm

Specifications

TR600 OTDR Optical Time Domain Reflectometry

TR600 Series meet the needs on the wide range of measurement by FTTx, Metropolitan Area Network, Backbone Network and other optical networks.

 TR600 OTDR Technical Specifications

Technical Specifications1

Wavelength (nm)

850/1300

1310/1550

Dynamic range (dB)2

23/21

32/30

38/36

Pulse width (ns)

5, 20, 40, 80, 160, 320, 640, 1280

5, 20, 40, 80, 160, 320, 640, 1280, 2560, 5120, 10240, 20480

Event blind zone (m)3

3m

3m

Attenuation blind zone (m)3

13m

13m

Linearity (dB/dB)

±0.05 dB/ dB

Loss threshold (dB)

0.05

Loss resolution ratio (dB)

0.01

Sampling resolution ratio (m)

0.125 to 8

Sampling point

32K

Distance uncertainty  (m)

±(1 m + 5×10-5× distance + sampling interval)

Distance scope (km)

0.3 to180

Typical real-time refreshing duration (s)

0.2

Memory capacity of trace

SD Card (4G), > 10000 pieces

Duration of measurement

Defined by user; 5sec, 10sec, 15sec, 30sec, 1min, 2min, and 3min are selectable

General specification

Dimension (H×W×D)

150×235×66

Relative humidity

0% to 95% (non condensation)

Weight

1.5kg

Temperature

Running temperature -10 to 50

Memory temperature -40 to 70

Power supply

Lithium battery; continuing working duration ≥ 8 hours

Warranty period

12 months

Remarks:

1. The technical specification describes the ensured performance of the instrument when using typical PC model connector to measure. Without considering the uncertainty caused by optical fiber refractivity.

2. Dynamic range is the data measured under the condition of the maximum pulse width and 3 minutes of average time. Dynamic range is the data measured under the condition of 180km/20480ns/3min.

3. Measuring conditions of blind zone: reflection event is within 4Km, reflection strength < -35dB. Measured by the minimum pulse width.

4. Dynamic range measured when there is filter.