Specifications

Specifications Parameters:

 

Test Items: Glucose(GLU),Bilirubin(BIL),specific gravity(SG),PH,Ketone(KET),occult blood (BLD),Protein(PRO),Urobilinogen (URO),Nitrite (NIT),Leucocyte (LEU),ascorbic acid(VC)

Working method: single-step,continuous

Test speed: 120 tests/hr(fast continuous),60 tests/hr(single-step, slow continuous)

Language Selection: Chinese, English (in Processing)

Repeatability: CV≤1%

Stability: CV≤1%

Recording Methods: Micro Heat-sensitive Printer,LCD Screen,Flash ROM

Screen: LCD Screen,resolution 192×64

Printer : Micro Heat-sensitive paper, row-width 57mm, Paper Roll diameter<50mm

USB x 2: Connected to external printer, barcode reader(Software in processing )

Communication port: RS-232C DB9 male 9600,19200,38400,57600, 115200bps Start 1 bit , data 8 bit; stop 1 bit, odd parity

Working temperature: 10℃-30℃

Relative humidity: ≤70%

Working environment: Stable horizontal table surface, fine grounding, avoid direct strong light source and strong electromagnetic interference source, etc.

Working Power: Wide power supply input AC100-240V 50/60Hz±1Hz

Fuse: 5mm×20mm,T2A/250V

Power Consumption of Input: 30VA

Safety Classification: According to Classification of Electric Shock Protection: Regular apparatus ,Class I Type B.