Product Main

Quick Details

Model Number: RM01
Brand Name: RAY MEDICAL
Place of Origin: China (Mainland)

Specifications

12.1" color TFT screen 
Multi‐channel display 
Big font interface
ST and arrhythmia analysis
Drug dose calculation and titration table
OxyCRG dynamic view display
Respiration monitoring with apnea alarms
Alarm review, arrhythmia review 
Patient info. Input 
Anti‐defibrillation design
NIBP over pressure protection
Plug‐in rechargeable battery, AC/DC available. 
Supporting wire or wireless network.
 
 
Technical Specifications
 
ECG
 
Lead Mode:  5 Leads (R, L, F, N, C or RA, LA, LL, RL,V) 
Lead selection : I, II, III, avR, avL, avF, V
Wave form :  2 ch
Lead mode: 3 Leads ( R, L, F or RA, LA, LL) 
Lead selection : I, II, III,
Wave form : 1 ch
Gain : 2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto
 
HR and Alarm Range :  Adult 15 ~ 300 bpm
                    Neo/Ped    15 ~ 350 bpm
Accuracy : ±1% or±1bpm, which is greater
Resolution : 1 bpm 
Sensitivity :  > 200 (uVp-p) 
Differential Input Impedance:  > 5 MΩ
 
CMRR:  Monitor > 105 dB 
        Operation > 105 dB 
        Diagnosis > 85 dB
 
Electrode offset potential: ±300mV 
Leakage Current : < 10 uA
Baseline Recovery : < 3 S After Defi. 
ECG Signal Range : ±8 m V ( Vp-p ) 
 
Bandwidth :  Surgery    1 ~ 20 Hz
            Monitor     0.5 ~ 40 Hz
            Diagnostic  0.05 ~ 130 Hz
 
Calibration Signal : 1 (mVp-p),  Accuracy:5%
ST Segment Monitoring Range : Measure and Alarm -2.0 ~ +2.0 mV
 
ARR Detecting  Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS, PNP, PNC,Alarm 
 
Available,Review Available
 
Respiration
 
Method:  Impedance between R-F(RA-LL)
Differential Input Impedance :  > 2.5 MΩ
Measuring Impedance Range :  0.3~3 Ω
Base line Impedance Range : 200~4 KΩ
Bandwidth : 0.1~2.5Hz
Resp. Rate :    Measuring and Alarm Range
              Adult 0 ~ 120 Brpm 
              Neo/Ped 0 ~ 150 Brpm 
              Resolution 1 Brpm 
              Accuracy ±2 Brpm
Apnea Alarm :  10 ~ 40 S
 
NIBP 
Method :  Oscillometric
Mode :  Manual, Auto, STAT                        
Measuring Interval in AUTO Mode : 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)
Measuring Period in STAT Mode : 5 Min
Pulse Rate Range : 40 ~ 240 bpm
Alarm Type : SYS, DIA, MEAN
Measuring and alarm range        Adult Mode:
                              SYS 40 ~ 270 mmHg 
                              DIA 10 ~ 215 mmHg 
                              MEAN 20 ~ 235 mmHg
                              Pediatric Mode
                              SYS 40 ~ 200 mmHg 
                              DIA 10 ~ 150 mmHg 
                              MEAN 20 ~ 165 mmHg
                              Neonatal Mode
                              SYS 40 ~ 135 mmHg 
                              DIA 10 ~ 100 mmHg 
                              MEAN 20 ~ 110 mmHg
Resolution : 1mmHg
 
Accuracy :  Maximum Mean error ±5mmHg
           Maximum Standard deviation ±8mmHg
Overpressure Protection:     Adult Mode 297±3 mmHg 
                         Pediatric Mode 240±3 mmHg 
                         Neonatal Mode 147±3 mmHg
 
SpO2
Measuring Range : 0 ~ 100 % 
Alarm Range :  0 ~ 100 % 
Resolution :   1 %
Accuracy : 70% ~ 100%  2 %
         0%  ~  69% unspecified
Actualization intervalabout : 1 Sec. 
Alarm Delay :  10 Sec.
Pulse Rate : Measuring and Alarm Range 20~300 bpm
           Resolution : 1bpm
           Accuracy: ±3bpm
                                   
TEMP
Channel :  2
Measuring and Alarm Range0 ~ 50 ℃
Resolution :  0.1 ℃
Accuracy: ±0.2 ℃
Actualization intervalabout :  1 Sec.
Average Time Constant :  < 10 Sec.
 
IBP
Label :  ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range : ART 0 ~ 300 mmHg 
                       PA -6 ~ 120 mmHg
                       CVP/RAP/LAP/ICP-10 ~ 40 mmHg
                       P1 / P2-10 ~ 300 mmHg 
 
Press Sensor :  Sensitivity 5  uV/V/mmHg
             Impedance 300-3000Ω
Resolution :  1 mmHg
Accuracy :  ±2% or 1mmHg which is greater
Actualization interval : about 1 secretary
 
Standard:  ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Optional:  Nellcor SpO2, EtCO 2, IBP, Thermal Recorder