Model Number: | 5E-AF4000 |
---|---|
Brand Name: | CKIC |
Place of Origin: | China (Mainland) |
Quick Details
Specifications
Standard configuration
1. Computer 2.Printer 3.Main analyzer 4.A/C adapter 5.Ash Cone Plate
6.Activated Carbon 7.Graphite 8.Dextrine 9.Magnesia 10.Combustion Cup
11.Duab kit 12.Ash cone module 13.Gasket kit 14.Tool kit
Optional configuration
7 samples ash cone plate (ISO std)
Features
High Automation
1.Determine four critical temperatures (DT, ST, HT, FT) automatically.
2. Rotary vertical sample-loader ensures smooth sample sending.
High Precision
1.2.0 Mega-pixel professional camera provides high resolution image.
2. Excellent image processing method and critical temperature measurement to ensure precise result.
3. Durable heating element with Si-Mo material and ergonomic vertical furnace design ensures stable working temperature.
4.Video clips available for replaying, which makes automatic identification and manually identification available.
Economic Operation
Atmosphere simulation by using graphite and activated carbon, minimize the operation cost without gas consumption.
High Safety
Automatically check the gas leakage with auditory alarm and gas flow terminated if triggered.
MODEL | 5E-AF4000 Ash Fusion Determinator |
Conforms to Method | ASTM1857, ISO540 and GB/T 219 |
Max. Sample loading | 5 Samples for GB and ASTM Standard, 7 Samples for ISO Standard |
Max. Temp.(℃) | Up to 1600 |
Temperature Control Resolution(℃) | ±1 |
Image Save Interval(℃) | 1 frame per 2 |
Heating Rate | Below 900℃ 15~30℃/min, Above 900℃ 5~10℃/min |
Analysis Atmosphere | Oxidizing Atmosphere / Reducing Atmosphere |
Power Supply | Single phase, AC220±10%, 50/60Hz, ≤2.6 kW |
Net Weight | 65kg |
Dimension(mm) | 480mm×470mm×740mm |