Place of Origin: | China (Mainland) |
---|---|
Model Number: | E6 |
Brand Name: | BAICHENG |
Type: | ECG machine |
Quick Details
Specifications
Features:
1. 1024*600 10inch
color touchscreen display
2. Twelve channel ECG machine with
interpretation
3. 12-lead ECG simultaneous acquisition
4. Multiple
operation modes: Manual/Auto/Arrhythmia Analysis Modes
5. 12-channel
recording with a rhythm lead
6. Digital filters eliminate EMG, AC, Baseline
Drift interference
7. Built in rechargeable Li-ion battery, 2 hours
continuous working
8. Adapt 110-240V, 50/60Hz AC power supply
9. 250
patients save/replay
10. Built in USB/RS232 interface
Power Supply |
100-120V or 220-240V, 50/60Hz or internal rechargeable battery with lifetime of 60 minutes |
Safety |
Class I, type CF |
LCD display |
Backlit high resolution 1024 x 600 10 inch touchscreen display 12 channel ECG waveform simultaneously |
Leads |
12 STANDARD or CABRERA leads |
Defibrillation protection |
Internal, according to current regulations |
Recording modes |
Manual, automatic sequential, automatic simultaneous, Rhythm lead selectable |
Print Channels |
12 channels , Rhythm lead selectable |
Writing system |
High-resolution thermal printer (8 dots/mm); Horizontal 40 dots/mm |
Thermal paper |
216mm wide 20M thermal paper in Z-fold packers or in rolls |
Paper speed |
5 - 6.25 - 10 - 12.5 - 25 - 50mm/s ± 3% |
Recording sensitivity |
Auto or Manual: 1.25 - 2.5 - 5 - 10 - 20 - 40mm/mV ±3% |
Input dynamics |
±500 mV @ 0 Hz ± 10 mV in the pass band |
Input impedance |
50 MΩ DC |
Power requirements |
AC220V/50Hz |
Common-mode rejection |
> 90 dB |
Frequency response |
0.05 to 150 Hz |
Time constant |
>3.3 s |
Data acquisition |
12 bits; 1000 samples/sec/channel when printing, filtering and computing |
Fuse Specification |
2Φ5x20mm |
Case storage |
150 cases at least, can be stored on external SD card |
Patient Cable |
standard 10 leads |
Resolution |
3.125 uV/bit |
Filters |
Low pass filter off 75 Hz, 100 Hz, 150 Hz |
Muscle filter: 25 or 35 Hz digital filters (-3 dB/oct) |
|
AC filter: digital notch filters 50 or 60 Hz selectable, - 20 dB or better |
|
Drift filter: 0.5 Hz high pass linear phase digital filter |
|
Net weight |
3.7kg (8.14 lb) |
Gross weight |
5.2kg (11.44lb) |