MSL 500ma x-ray machine MSLHX06 Production Specification
High frequency and pressure generator: Size: 600*420*650mm
the high frequency multiple-pulse X ray generator is used for radio diagnosis, it ensures data and repetition's accuracy and shorten time of exposure.
with the most professional and stable high frequency and pressure generator, it makes a high dose output,high power load and ultrahigh frequency contravariant which perfectly guarantees users' radiography quality. The generator is compatible with almost all good brands of tube in the world.
it is equipped with APR/AEC which possess tube overload protection)
the control functions of the generator are integrated in the system control center
It can meet high frequency filming requirements since its rated voltage is with a high of 150KV
50KW large capacity makes a longer uninterrupted work
perfect performance and stability
it is complicated to set parameters for most of generators which lead to time wast. But our generator is equipped with easier parameters settings and firmwares updating programs.
rotating anode's calorific capacity display and protection can prevent overheat of pipe shroud cause by overload.
It stops the rotating anode by DC, it makes less noise after exposure.
X Ray tube module
1.module is rated voltage is 150KV,it guarantees stabilization of operation.
2.Focus: small focus: 0.6, big focus:2.0, focus power is high enough to make operation stable.
3.X ray module is equipped with heat protection switch.
column rack
1.new generation light-design saves much space. easy to assemble.
2.It is convenient for clinical application, satisfying multi-dimension filming requirements for every part of human body.
3.flat bed can only be moved when it is being charged.
U type filming bed
Dimension(length*width*height):
2200mm×750mm×660mm
Dimension(length*width*height):
Lengthways moving range: 1000mm
Transverse moving range: 230mm
Moving range of flat panel detector holder: 350mm
Load-bearing of the filming bed: no less than 150kg
Thickness of filming bed board: 8mm solid carbon fiber plate
height of vertical
1.height of vertical type filming racks: 1920mm
2.Up-and-down motion range of flat panel detector:1300mm( manual operation)
Specifications of tube support:
Stand column height: 2000mm( including ground rack)
Size of ground rack: 3100*460*40mm
X ray tube holder's up-and-down moving distance: 1250mm(manual operation)
X ray tube holder's left-and-right moving distance: 2000mm( manual operation)
X ray tube's angle of rotation range: 0°±2°~ ±135°±2°
Arm stretch length of X ray tube arm: 200mm
X ray tube stand column rotation range: ±360°
Digital Filming machine MSLHX06 for sale Packing list
NO. | nae | unit |
1. | X ray tube holder | one set |
2. | sternum stand | one set |
3. | filming bed | one set |
4. | X ray tube | one set |
5. | Beam-defining clipper | one set |
6. | high frequency and pressure X ray generator for medical diagnostic use | one set |
7. | operating console | one set |
8. | high voltage cable | two faculties |
Configuration parameter of the x-ray machine
NO. | technical requirements |
1 | high frequency and pressure X ray generator for medical diagnostic use |
1.1 | input power: 380V±38V, 50Hz±1Hz. |
1.2 | maximum power output :50 kW |
1.3 | Photography tube voltage:it can be adjusted continuously, adjustable step width is 1KV,adjustable range: 40kV~150kV |
1.4 | photography tube current:it can be adjusted by sections. Adjustable range: 10mA~630mA |
1.5 | load time: adjustable range: 0.001-10sec |
1.6 | current can be adjusted by sections. Adjustable range: 0.1mAs~500mAs |
1.7 | working frequency: 100kHz |
2 | X ray tube |
2.1 | : highest peak tube voltage: 125kV |
2.2 | maximum filament current's heat characteristic: big focus:5.4A, small focus:5.4A. |
2.3 | focus: small focus:0.6mm;big focus:2.0mm |
2.4 | Anode input power:small focus:18kw;big focus:50kw |
2.5 | Anode optics angle: 14 |
2.6 | rotate speed of Anode: 2800rmb |
2.7 | Anode thermal capacity: 160kJ (220kHU) |
2.8 | tube module thermal capacity: 1250KHU(900KJ) |
3 | Filming bed |
3.1 | length: 2200mm±10% |
3.2 | width: 750mm±10% |
3.3 | lengthways moving range of bed surface: 1000mm±10% |
3.4 | crosswise moving range of bed surface: 250mm±10% |
3.5 | bed surface locking pattern: electromagnetic lock(permanent magnet) |
3.6 | bed surface locking control: foot switch |
3.7 | breaking force: ≥100N |
3.8 | starting force:≤50N |
3.9 | load bearing of bed panel: ≥150kg |
4 | stand column |
4.1 | height: 1920mm±10% |
4.2 | up-and-down moving range: 420mm-1810mm±10% |
4.3 | up-and-down moving range along the lathe bed: 2000mm±50mm |
4.4 | vertical lift range along the lathe bed: 430mm-1660mm±10% |
4.5 | left and right moving range of flat panel detector : 350mm±5% |
4.6 | X ray tube module's slewing range around the cross arm(power brake locking): 360° |
4.7 | X ray tube slewing range: 0° ±2° ~ ±90° ±2° |
4.8 | stand column slewing range: fixed |
5 | beam-defining clipper |
5.1 | power supply: 24VAC,7A |
5.2 | inherent filtration : 2.0mmAL/75KVp |
5.3 | it is easy to turn on and off, flexible and reliable. It makes different sizes of rectangles exposure field in accordance with various requirements in practice. |
6 | others |
6.1 | high voltage cable:≥8m |