Product Main

Specifications

Characteristics: 

Completely controlled by PC, can flexibly select the flame and graphite furnace atomizer (optional)

Integrated floated optical platform design can obviously improve the optical system shock resistance and can keep stable though use the optical signal for a long term. 

Eight light stands can be changed automatically and preheat the eight element lights meantime as well as optimize the working condition of the hollow cathode lamp

Position adjusting: the best height of the flame burner and can automatically set the front and rear positions

Fully automated wavelength scanning and peak searching

Complete safety chains protection equipment: the function of warning and automatic safety protection towards the wrong burner, leakage of the gas, under voltage of air and the abnormal flameout

Deuterium lamp and self-absorption background regulation 

Data processing: super strong database, possesses more than 500 data self-storage and cut-off storage function, can store the analyzed result with the form of EXCEL and the testing method and the result can be randomly called

Measuring method: flame absorption method and emission method

Result printing: parameter printing, data result printing and diagram printing

 

Complete automatic controlling system: 

With the help of the software, the following can be easily achieved

Selection of the element lamp

Up-down-front-rear adjustment of the lifter

Adjustment of the optical energy

Selection of the slit

Determination of wavelength scanning and peak searching

Selection of the atomizer

Setting of the background deduction method 

Controlling of the gas flow

Automatic flaming and flaming out 

Setting of the graphite furnace testing method 

 

 

Technical Specifications of  4530F Minerals Testing AAS Machine

Wavelength range:190 ~ 900nm

Spectral bandwidth:Can automatically change between 0.1nm, 0.2nm, 0.4nm, 1.0nm and 2.0nm

Accuracy of the wavelength:±0.15nm

Receptivity of the wavelength≤0.04nm

Base line stability:≤0.002A/30min (Cu)

Characteristic viscosity:0.02μg/ml/1% (Cu)

Checking limit:0.004μg/ml (Cu)

Preciseness:0.5%

Grating:1800 lines/mm

Inflamer:All-metal titanium burner

Atomizer:Effective glass atomizer

Lamp stand:Eight lamps 

Dbackground correction ability:When the background is 1 A, the background ability should be deducted not less than 50 times; self absorption background deduction method

Size and weight:700mm×550mm×450mm  75Kg

Power source:220V±22V  AC

  

Comparison of AAS320N Atomic Absorption Spectrophotometer and  4530F Minerals Testing AAS Machine

Model

AAS320N

AAS4530F

Wavelength range 

190 – 900nm

190 – 900nm

Wavelength accuracy 

±0.5nm

≤±0.15nm

Wavelength repeatability 

0.3nm

0.04nm

Spectrum bandwidth 

0.2,0.4,0.7,1.4,2.4,5.0 nm

0.1,0.2,0.4,1.0,2.0nm

Base line stability

≤0.004A/30min (Cu)

≤0.002A/30min (Cu)

Characteristic viscosity

0.04μg/ml/1% (Cu)

0.02μg/ml/1% (Cu)

Checking limit

0.008μg/ml (Cu)

0.004μg/ml (Cu)

Background calibration  ability

>30 times

>50 times