Product Main

Quick Details

Wavelength Range: 0.1 nm
Weight: 55 lbs
Model Number: LAMBDA 365
Brand Name: Perkin Elmer
Place of Origin: Malaysia

Specifications

Specifications:
 

Parameter LAMBDA 365
Source Lamp Tungsten/Deuterium
Monochromator Blazed holographic grating with 1200 gr/mm
Czerny-Turner with 0.2 m focal length
Detector Si diodes, Sample and Reference
Photometric System Double Beam Optics
Wavelength Range 190-1100 nm
Spectral Bandwidth 0.5, 1, 2, 5, 20 nm
Data Interval 0.05, 0.1, 0.5, 1, 2 nm
Wavelength Accuracy ± 0.1 nm at D2 peak  656.1 nm, ± 0.3 nm for full range
Wavelength Reproducibility ± 0.1 nm
Photometric Range ± 4A
Photometric Accuracy ± 0.002A (0 - 0.5 Abs)
± 0.004A (0.5 - 1 Abs)
Photometric Reproducibility < 0.001 AU  0.5 and 1.0 AU
Stray Light < 1.0% (198 nm KCL)
< 0.02% (220 nm Nal)
< 0.02% (340 nm NaNO2)
Baseline Stability < 0.0003 Abs/h (700nm, after one hour warm up)
Baseline Flatness ± 0.00005 Abs (after one hour warm up)
Photometric Noise < 0.00005 Abs (at 700nm, RMS)
Photometric Stability N/A
Maximum Scan Rate 3000 nm/min
Maximum Slew Rate 8000 nm/min
Interface USB
Weight 55.2 lbs. (25 Kg)
Dimensions 19.5"W x 19.7"D x 10.6"H (495 x 500 x 270 mm)
Electrical 100-240 V 50/60 Hz
Environmental Requirements Temperature: 15-35°C
Humidity: 30 - 80%
Non-condensing
  • For UV/Vis Spectrophotometric analysis of laboratory samples
  • Double beam optics for high stability and low stray light
  • Variable spectral bandwidth from 0.5 nm to 20 nm
  • Extremely high signal to noise ratio
  • Modular design and a large number of accessories for maximum usability
  • Large sample compartment fits more than 10 sampling accessory combinations
  • Optional security software for 21 CFR part 11 compliance
  • Ideal for use in biotechnology, environmental, genetics, manufacturing, pharmaceutical QA/QC and academic labs