Usage: | Electric Safety Test |
---|---|
Power: | Electronic |
Model Number: | PG-9596 |
Brand Name: | PEGO |
Place of Origin: | China (Mainland) |
Quick Details
Specifications
Electric Safety Comprehensive Test System (6 in 1)
Features
Test function: AC withstand voltage test, insulation resistance test, leakage current test, ground resistance test, power test and voltage start up test.
Meet the requirements of IEC60335-1, IEC60950, IEC60598-1, IEC60745-1 and IEC60990.
Applied to household appliance, motor, electric tool and other manufacturing industries, Auto switch test functions, decrease the cost.
Test accuracy: 2%
Supper-big LCD display (English version), good human-machine interface.
PLC control, RS232/RS485 interface port
Preset warning function, system self-checking function, auto protection function, memory analysis function
Specification
Input |
||
Input voltage |
220Vac±10% |
|
Input frequency |
50/60Hz ± 5% |
|
Fuse |
Fuse 5A/250V |
|
AC Withstand Voltage Test |
||
Rated output load |
100VA(AC5000V/20mA) |
|
Range of output voltage and accuracy |
300-5000V , ± (2% setting value + 3V) |
|
Output frequency and accuracy |
50Hz/60Hz, ±0.2% |
|
Output waveform |
Sine wave, crest factor= 1.3 - 1.5 |
|
Output stability |
± (1% output value + 5V), no load to full load |
|
Range of output current and accuracy |
0.10-20.00mA, ± (2% setting value + 0.03mA) |
|
compensation current range and accuracy |
0.0-1.00mA ± (2% setting value + 0.03mA) |
|
Test time and resolution |
1-999s, resolution: 1s |
|
Insulation resistance test |
||
Rated output load |
2VA(DC1000V/2mA) |
|
Range of output voltage and accuracy |
DC300-1000V ± (2% setting value+ 3V) |
|
Output voltage ripple factor |
<1%(pure resistance 1000V/1mA) |
|
Test range |
1.0-10.0-1000MΩ |
|
Accuracy |
(100MΩ ±(5%+3MΩ) , >100MΩ ±(10%+5MΩ) |
|
Test time and resolution |
1-999s, resolution: 1s |
|
Ground resistance test |
||
No-load output voltage |
<8VAC |
|
Range of output current and accuracy |
5-30A ± (2% setting value) |
|
Output current stability |
0.3%(10A/100mΩ) |
|
Range of output frequency and accuracy |
50Hz/60Hz ±0.2% |
|
Test range and accuracy |
5-600mΩ(5-10A); 2 – 200mΩ (11-25A); 2–150 mΩ(26-30A); ± (2% setting value + 3mΩ) |
|
Test time and resolution |
1-999s, resolution: 1s |
|
Leakage current test |
||
Range of output voltage |
1.06 times of input voltage, Pressure regulation 5% |
|
Range of output voltage and accuracy |
10-255VAC, ± (2% reading value + 3V) |
|
Test range of current and accuracy |
0.050~10.00mA, ±(2% reading value+0.030mA) |
|
compensating current range and accuracy |
0-0.100mA , ± (2% setting value+0.030mA) |
|
Test time and resolution |
1-999s, resolution: 1s |
|
Test mode setting |
dynamic/static |
|
Range of output power |
2KW/5KW |
|
Power test |
||
Range of output voltage |
1.0 times of input voltage, Pressure regulation 5% |
|
Range of output voltage and accuracy |
10-255VAC ± (2% reading value+ 3V) |
|
Range of current and accuracy |
0.10A~20.00A ±(reading value×1.0%+range×1.0%) |
|
Range of power |
10~6000W ±(reading value×1.0%+range×1.0%) |
|
Test time and resolution |
1-999s, resolution: 1s |
|
Range of output power |
2KW/5KW |
|
Start up test |
||
Range of output voltage |
0.85 times of input voltage, Pressure regulation 5% |
|
Test Range of output voltage and accuracy |
10-255VAC ± (2% reading value+ 3V) |
|
Test range of output current and accuracy |
0.10A~20.00A ±reading value×1.0%+range×1.0%) |
|
Test time and resolution |
1-999s, resolution: 1s |
|
Range of output power |
2KW/5KW |
|
Characterisitcs |
||
PLC control device |
Signal input: |
Start, stop, |
Signal output: |
Alarm lamp |
|
Memory group |
15 groups, 7 steps for each group |
|
Dsiplay |
240*128 LCD display |
|
Interface port |
RS232/RS485 |
|
Correction mode |
By front panel or software |
|
Working condition |
0-40℃, 20-80%RH |
|
Dimension |
470mm(W)×1000mm(H)×520 mm(D) |