Place of Origin: | China (Mainland) |
---|---|
Model Number: | TOP |
Brand Name: | DOTOP |
Quick Details
Specifications
Patient Monitor
Standard parameter: ECG, RESP, NIBP, SpO2, TEMP, PR.
12.1" large-screen color TFT display
Built-in high-capacity battery, up to 4 hours working time
7-lead ECG waveforms display ,big fonts observational interface
Reviewable ECG waveforms, Maximum 240 hours reviewing graphic trends
Digital SpO2 technique, precise measurement for weak perfusion and finger shaking
Arrhythmia and S-T segment analysis and pacemaker detection
Protection against interference from defibrillation and electrosurgical Unit
Applicable to ICU, CCU, operation, adult, pediatrics and neonatal patient
ECGl Full lead: I, II, III, AVL, AVR, AVF and V
l Gain 5mm/mV, 10mm/mV, 15mm/mV, 20mm/mV and 25mm/mV
l Scanning speed: Changeable with three grades of 12.5mm/s, 25mm/s and 50mm/s
l Band width 0.05Hz-100Hz(3dB)
l Heart rate:
Scope: 15 times/minute to 300 times/minutes
Accuracy: ±1 time/minute
l Variation of heart beat LORENZE drawing: Analysis on heart beat data for the current 5 minutes
with histogram during RR period.
l Arrhythmia :Twenty kinds of common abnormal ECG can be analyzed, with ECG data of 1000 times
of arrhythmia kept and with 10 seconds kept for each time.
Accuracy: Error rate < 20%
Undetected rate < 20%
Analysis function should be started when using the above two kinds of function.
l ST section
Scope: ±0.8mV
Precision : <±0.05mV
SpO2l Detection scope: 0%-100%
Accuracy ±1% 90%-100%
±2% 70%-89%
NIBPl Theory Electric oscillator
l Cuff gassing time less than 15 seconds
l First-time gassing pressure: dynamic state 24kPa (180 mmHg)
Manual selection: 9.3kPa (70 mmHg), 13.3kPa (100 mmHg), 16.0kPa
(120 mmHg)
18.6kPa (140 mmHg), 20.0kPa (150 mmHg), 21.3kPa (160 mmHg),
l Automatic detection period: 1 minute to 240 minutes
Stop size: 1 min(1 min -10min), 5min(10 min -30min), 10min(30 min -90min) and 30min(90 min -240min)
l Detection scope:
Systolic blood pressure: 4kPa -34kPa(30mmHg -255mmHg)
Diastolic pressure 2kPa -29.3kPa(15mmHg -220mmHg)
Mean pressure 2.7kPa~31.3kPa (20mmHg~235mmHg)
l Detection accuracy
Static pressure: ±3mmHg
Pulse rate: ±2%
Respl Chest impedance method or concentration of carbon dioxide at end of respiration
l Scope of chest impedance method 0 time/minute – 100 minutes/minute
l Precision of chest impedance method 1 time/minute
l Detection scope of carbon oxygen at end of respiration 3mmHg-75mmHg; 0.4%vol-9.9%vol
l Detection scope of respiration rate: 2 times/minute ->60 times/minute (including baby)
l Detection precision of respiration rate: ±1 time/minute
Templ Single/double path
l Type: Vascular/surface temperature
l Scope: 20℃~50℃
l Accuracy ±0.2℃
Invasive blood pressure (selected mating)l Single/double path
l Type: direct and continuous detection
Scope: -6.67kPa(-50mmHg)—46.67kPa(350mmHg)
l Precision: below 13.33kPa(100mmHg) ±0.27kPa(2mmHg) 13.33kPa(100mmHg) above ±2%
l Frequency response: DC-12Hz
l Input resistance: not less than 500kΩ
l Sensitivity of sensor: 5μV/V/mmHg
l Motivation voltage: 5V (DC)
l Accuracy of sensor: ±2%
l Isolation voltage: 4 kV