Product Main

Quick Details

Dimensions: 285mm85mm56mm
Measuring Resistance Range: 0.01-1200ohm
Measuring Current Range: 0.00mA-20.0A
Model Number: MEWOI3100D+
Brand Name: MEWOI
Place of Origin: China (Mainland)

Specifications

INTRODUCTION

MEWOI3100D+ Clamp Earth Resistance Tester is widely applied in ground resistance measurement, loop resistance measurement in fields such as electric power, telecommunications, meteorology, oil field, architecture and industrial electrical equipment, etc. When measuring ground system with loop, there is no need to disconnect the ground wire without auxiliary electrode, which means safer and faster. MEWOI3100D+ Clamp Earth Resistance Tester can measure the ground fault which can not be measured by traditional method, can apply in the occasion where traditional methods are unworkable; MEWOI3100D+ Clamp Earth Resistance Tester measures the integrated value of ground resistance and resistance lead. What's more, MEWOI3100D+ Series tester can measure the leakage current of ground wire.


Technical Specification

Function

Ground resistance test, Loop resistance test

Power Supply

6VDC 4 PCS of AA Alkaline batteries

Resistance Range

0.01-1200Ω

Resistance Resolution

0.001Ω

Highest Resistance Accuracy

±1%±0.01Ω

Current Range

0.00mA-20.0A

Current Resolution

0.05mA

Highest Current Accuracy

±2.5%±1mA

Data memory

99 Sets

Alarm Threshold Setting Range

Resistance: 1-199Ω

Alarm Sound and Light

" Beep-beep-beep" alarm tone, Alarm Symbol Flashing, Press AL key to turn on, turn off Alarm or set Alarm Threshold.

Interference Signal Recognition

 

" Beep-beep-beep" tone, NOISE symbol blinking

Working Temperature and Humidity

-20℃-55℃; 20%RH-90%RH

LCD Screen

4 Digits LCD display, Screen L×W: 47mm×28.5mm

Clamp Size

Φ32mm

Clamp Opening Extent

32mm

Tester Weight

1160g (With batteries)

Clamp Dimension

L×W×T: 285mm×85mm×56mm

Protection Level

Dual Insulation

Structural Feature

Clamp CT

Shifting

Automatic shifting

External magnetic field

<40A/m

External electrical field

<1V/m

Single measurement time

0.5 second

Resistance measurement frequency

>1KHz

Measured current frequency

50/60Hz Automatic measurement

Low Battery Indication

Low battery symbol displays when battery is lower then 5.3V

Memory Full Indication

MEM symbol blinks when all 99 sets memory is full

Over Range Indication

Out of range indication " OLΩ" , " L0.01" or " OL A"

Auto Power-off

5 minutes after Boot up

Power Consumption

Max 50mA either in booting up or normally working, Continuously working for 30 hours

RS232 Interface (Optional)

The data stored in the meter are uploaded to the computer through RS232 interface and data wire, Then generate report files

Safety Provisions

IEC1010-1, IEC1010-2-032, Pollution level 2, CAT Ⅲ(150V)

Accessories

Meter: 1 piece, Test ring: 1 Piece, Meter Suitcase: 1 Piece


SELECTION GUIDE

Digital Earth Resistance Clamp Meter

Click "blue"marked model No. for detailed Specificaiton

Model Range
Resolution
clamp size RS232 Data Storge Audible and visible alarm  Other
MEWOI3000B+ 0.01Ω-200Ω     0.01Ω 65mm×32mm optional 
99sets
MEWOI3100B+ ?32mm
MEWOI3000+ 0.01Ω-1200Ω    0.01Ω 65mm×32mm optional  99sets
MEWOI3100+ ?32mm
MEWOI3000D+

0.01Ω-1200Ω 

0.00-20.00A  

0.01Ω,0.05mA   65mm×32mm optional  99sets
MEWOI3100D+ ?32mm
MEWOI3800C+ 0.01Ω-200Ω     0.01Ω 60mm*30mm Appoint with RS485 communication interface, provide standard MODBUS-RTU communication protocol ×

MEWOI3800D 0.01Ω-200Ω   0.01Ω 60mm*30mm ×
LCD display,warning function.
MEWOI4000

0.01Ω~2000Ω;

 0~600V

0.01Ω,1V
400sets
2/3-wires measurement of earth resistance/voltage
MEWOI4000C

earth resistance:0.01Ω~30.00KΩ;

soil resistivity:0.01Ωm~9000KΩm;

 0~600V

0.01Ω,0.01Ωm,1V
300sets
2/3/4-wires measurement of earth resistance,soil resistivity,and earth voltage
MEWOI4200

earth resistance:0.01Ω~30.00KΩ;AC Leakage current: 0.00mA~600A; 

soil resistivity:0.01Ωm~9000KΩm;Earth Voltage:0.0~100.0V,dual clamp: 0.01Ω-100Ω

0.01Ω,0.01Ωm,0.01A,0.1V ?68mm 2000sets
2/3/4-wires,selection & dual clamp method to measure the earth resistance,soil resistivity,earth voltage,AC voltage,AC leakage current,warning function.