Product Main

Specifications

AX31 is used for First-Aid, Ambulance, Emergency situation and patients transport

in hospital.

It can be used for child and adult patients.

Features:

Multiple functions, compact design, easy to take, design for transport and first aid.

The main parts adopt good-quality components, accurate and reliable.

LCD screen, simple and intuitive operation.

3 kinds of power sources can be chosen: AC, DC and Internal Battery.

 

Environmental Requirements

 

Relative humidity: 15%~95%

Ambient temperature:  -18~ 50°C

Atmospheric pressure: 70~110kPa

Gas source: 0.28~0.6Mpa 60L/min

Power: AC100V-240V or DC12V

Li-battery:  2200mAH, rechargeable

 

 

Classification

 

Class, Type B—by protection against electric shock.

IPX4—by protection against harmful ingress of water, particulate matter.

It is suitable for the patient whose weight is more than 25KG

 

Configuration

 

Standard Configuration: Ventilator, Oxygen Pressure Reducer, High-Pressure oxygen hose, Lung Simulator, Breath Circuit, Mask, Head Band, Power Adapter.

Optional: Carrying Package, 2L oxygen Aluminum Cylinder, Oxygen Bridge, Hanging Rack.

 

Modes:   CMV, A/C, SIMV, SPONT

Tidal Volume: 50~1200mL

Respiratory Rate:  5~60bpm,1-12 bpm for SIMV

FiO2: 40%~100%

Trigger: -10~0hPa

Max Minute Volume:≥22L/min

PEEP: 0, 5 ~20hPa

PIP: 0~60hPa

Over Pressure Relief: ≤ 72hPa

Sigh:      0~10 per 100 breath

Base Flow: 0~5 L/MIN

Package Dimension: 530*530*280mm

Weight: 13kg(Including all the accessories)

 

Monitors:      Tidal Volume; Sigh; Respiratory Rate; Peak Airway Pressure; Trigger; Airway Pressure; Airway Pressure Wave Form

 

Alarms: High Airway Pressure Alarm; Low Airway Pressure Alarm; Low Battery Alarm