Place of Origin: | China (Mainland) |
---|---|
Brand Name: | Sonoicare |
Quick Details
Specifications
Specification
Sample type: Anti-coagulation whole blood, serum, plasma.
Sample volume: 100ul.
Bar code: Two-dimensional bar code.
Testing time :12mini/person
Testing principle : absorption spectroscopy, transmission turbidimetry
Testing method: end point, kinetic, fixed time, turbidmetry etc.
Temperature: 37℃±0.2℃.
Absorbance :0-4.5AbsResolution: 0.001Abs.
Carryover: 0
QC & Calibrate: auto test in real time.
Work environment: Temperature:10-32℃ Humidity:<90%
Optic system: Reversed optic. 340nm, 405nm, 450m, 505nm, 546nm, 578nm, 630nm, 850nm.
Light source: 12V/20W,.Tungsten halogen lamp`s life span reaches over 2500hours.
Power supply:AC100V-240V,50-60Hz;Power: 70W
Interface:7.0 inch True color LCD touch screen,multi-language choice.
Vet version is optional.
Storage: >100000 results.
Printer: built-in thermal printer and external printer.
Data port: 4 USB, 1 RS232. Software upgrade function by WIFI.
Weight: 5KG
Carton Size: 775px(H)*525px(W)*28(L)cm
Delivery Detail: within 3 days after payment
FAQ
Q: What are the power requirements to operate this analyzer?
A: The analyzer requires 100-240 volts AC, 50-60 Hz.
Q: What is the shelf life of the reagent discs?
A: The reagent discs can be stored for up to 12 months at 2-8°C.
Q: Can the expired reagent discs be used?
A: No, the result of the test will be unreliable.
Q: How long can the reagent disc be used after removing from the foil pouch?
A: After opening the pouch, the disc should be used within 20 minutes. Do not place the disc back into the refrigerator for later use.
Q: Can reagent discs be used again?
A: No, it can only be used once. It is a single-use disposable reagent disc. One disc one sample.
Q: What types of samples can be used with this analyzer?
A: The sample can be whole blood, plasma or serum.
Q: What is the sample volume required for a test?
A: 100ul is required to run a test. In case of rechecking, it is recommended that collecting blood sample is not less than 250ul.
Q: Before testing, does the reagent disc need to take out from refrigerator to guarantee the test temperature?
A: Yes, to avoid congealing and guarantee test temperature, before unpack and test, reagent disc, have to be placed 20 minutes after taking our from refrigerator or low temperature storage Tank.
Q: In the process of sample and control test, when clicking STOP button, can the test disc be used as a new disc again?
A: The stopped test disc can`t be used again. If it is not must situation, please don`t stop it.
Q.Do the users need to add reagent?
A: No, the reagent is already packed well in each of the disc hole before packing foil pouch.
Q:How to maintain the instrument?
A:Keep the instruments shell and drawer clean. Compared to the traditional analyzer, our instrument has less maintenance.