Classification: | Other |
---|---|
Place of Origin: | China (Mainland) |
Model Number: | YKG-800A |
Brand Name: | YLK |
Quick Details
Specifications
YKG Air Purification System Glove Box is a kind of laboratory equipment which can provide the sealed environment for sample operation and testing. By the automatic air purification system, the Glove Box system will cycle to remove internal active substances. It allows the system always maintain clean and high purity inert gas environment, and keep the water & oxygen contents in operation box to 1ppm below.
Gas Pressure Withstand Range | -3000Pa ~ +3000Pa |
Water Contents Under Standard Case ( 1 Pa, 20 degree ) |
≤1ppm |
Oxygen Contents Under Standard Case ( 1 Pa, 20 degree ) |
≤1ppm |
Leakage Rate ( 1 Pa, 20 degree) | ≤0.05Vol%/h |
GLOVE BOX APPLICATION
- Experiment which needs anhydrous, anaerobic and ultra-clean work environment.
- The R&D and production of batteries and battery materials.
- The R&D and manufacturing of special lamps:HID lamps, metal halide,ceramic metal halide.
- Welding: resistance welding, TIG welding, laser welding, plasma welding, brazing
- The development and production of medical supplies.
- Development and production of the super capacitor.
- Fine chemicals, nuclear industry.
- The new energy and new materials development and production.
SYSTEM INTRODUCE
1. Sealed Cycle System:
It forms a sealed cycle system by vacuum glove box and air purification system. And the air purification system will provide the power of air cycle.
2. Interface:
Equipped with convenient colorful touch screen with Chinese / English / Russian language interface.
3. The Control Systems:
Pressure Control System: automatic/by manual
Air Purification System: automatic/by manual
Air Purification Restore System: automatic
Box Inner Condition Alarm System: automatic
Box Inner Condition & pressure Parameter Set: automatic/by manual
Parameter Set of Automatic Air Purification System
Look up the history records of system;
Detect the working condition of Air Purification System
Alarm System for operation errors and warns
4. Reduction & Reproduction:
The Purification System applies the reproducable & reducible purification material by automatic system control. To achieve the reduction & reproduction, it should connect the power and be supplied with hydrogen(5-10%) with inactive gas mixed.
5. Mis-operation Avoid:
Apply the PLC program control system to detect the working condition of environment & machine. And the system will indicate with the Operation Error or Condition Inadequate alarm if there's any problems to ensure the safety & convenience.
MODELS FOR CHOOSE
Model |
Gloves (pcs) |
Operation Side |
Purification Column (pcs) |
Operation Box Size |
Overall Dimension |
YKG-800A | 2 | 1 | 1 |
1200*800 *930mm |
1915*830 *1830mm |
YKG-800D | 2 | 1 | 2 | ||
YKG-1000A | 4 | 2 | 1 |
1200*1000 *930mm |
1915*1060 *1830mm |
YKG-1000D | 4 | 2 | 2 | ||
YKG-1200A | 4 | 2 | 1 |
1200*1200 *930mm |
1915*1260 *1830mm |
YKG-1200D | 4 | 2 | 2 |
Gloves Number: 2/3/6/8 pcs for Choose
Operation Side: 1/2 sides for Choose
Purification Column: 1/2 pcs for Choose
Size of Operation Box:
1200*800*930mm, 1200*1000*930mm, 1200*1200*930mm;
1500*800*930mm, 1500*1000*930mm, 1500*1200*930mm;
1900*800*930mm, 1900*1000*930mm, 1900*1200*930mm;
2350*800*930mm, 2350*1000*930mm, 2350*1200*930mm,