Usage: | thermal shock test chamber |
---|---|
Power: | Electronic |
Model Number: | BTS Series |
Brand Name: | DGBELL |
Place of Origin: | China (Mainland) |
Quick Details
Specifications
Applications
This thermal shock environmental chamber (cold and hot air thermal shock test chamber) is widely applied forelectronic and electric products, components and materials by constant high low temperature shock and rapid temperature change reliability test.
Standard compliance1.Manufacturing Standard:
GB/T10586-2006 Specifications for damp heat chambers
GB/T10592-1989 Specification for low/high temperature test chambers
2.Inspection Standard
GB/T5170.2-1996 Inspection methods for basic parameters of environmental testing equipments for electric and electronic products. Temperature testing equipments;
GB/T5170.5-1996 Inspection on methods for basic parameters of environmental testing equipments for electric and electronic products--Damp heat testing equipments
3.Testing Standard Compliance:
GB2423.1-2008(IEC68-2-1)Environmental testing - Part 2: Test methods - Tests A: Cold
GB2423.2-2008(IEC68-2-2)Environmental testing for electric and electronic products Test methods - Tests B:Dry Heat
GB/T2423.22-2008 Environmental testing for electric and electronic products-Part 2:Test methods-Test N:Change of temperature
GJ8150.3A-2009(MIL-STD-810D)environmental test methods and engineering guidelines
GJ8150.4A-2009(MIL-STD-810D)environmental test methods and engineering guidelines
GJB/150.5-2009 Temperature Shock Test
Specifications
Model No. |
Internal Dimension (mm) |
External Dimension (mm) |
BTS-50 |
W350*H400*D360 |
W*1320*H1860*D1420 |
BTS-80 |
W400*H500*D400 |
W1320*H2180*D1520 |
BTS-150 |
W500*H600*D500 |
W1450*H2200*D1700 |
BTS-250 |
W600*H700*D600 |
W1550*H2280*D1800 |
BTS-512 |
W800*H800*D800 |
W1800*H2300*D2000 |
BTS-1000 |
W1000*H1000*D1000 |
W2000*H2350*D2300 |
Technical Parameters
Pre-cooling Zone Temp. |
-10~-80℃ |
Pre-heating Zone Temp. |
+50~+220℃ |
Pre-cooling Zone Cooling Rate |
-10~-40℃<30 mins -10~-60℃<50 mins -10~-80℃<65 mins |
Pre-heating Zone Heating Rate |
+50~150℃<30 mins +50~180℃<50 mins +50~220℃<60 mins |
Temp. Shock Range |
B type :-40~150℃ C Type :-55~150℃ D Type :-65~150℃ |
Temp. Recovery Time |
<5mins |
High low Temp. Exposure Time |
<30 mins |
Temp. Fluctuation |
≤±0.5℃ |
Temp. Deviation |
≤±2℃ |
Temp. Uniformity |
≤±2℃ |
Features
With high precision perfect external design, external with double sides cold rolled plate electrostatic powder coated material, internal with SUS#304 high temperature resistant stainless steel. Insulation material adopt fire resistant high strength PU polyurethane foaming thermal insulating material. More than 20% energy saving is achieved & thanks to the introducing of Japan & German advanced refrigerator control technology. Control system and control circuit components are introduced with famous brand.