Specifications

Gas chromatographygas analyzerlaboratory instrument

·Features

1. the temperature control accuracy can be precise to 0.01 oC by the new integrated digital electronic circuit with high control precision and reliable performance.
 
2. It has the functions of self-diagnosis for boot, stopwatch (for determining the flow), timer operation, storage protection for power off, keyboard locking, anti-interference, network data communication and remote control.
 
3. It can gets lower limit of detection by unique injector system design.
 
4. You can choose automatic/manual gas six-way valve injector, headspace injector, thermal desorption injector, cracking furnace injector, methane conversion furnace and auto-injector.
 
5. the program heating and the furnace temperature precision can be controlled stable and fast.
 
6. Optional detectors are FID, TCD, ECD, FPD, NPD and PDHID. All detectors can control the temperature individually. FID is easy to assemble and disassemble for cleaning or changing the nozzle. TCD stable time is ≤20min.
·Specification
Column Oven
Furnace dimensions: 28 x 30 x 18cm
Temperature operating range Higher than the room temperature of 5oC-450oC
Temperature setting 1oC;programmed heating rate 0.1oC
Maximum heating speed 120oC/min
Temperature stability It will be 0.01oC, when the environment temperature changes 1oC.
Programmed temperature steps 7 steps 
Column loss can be compensated (dual channel column)
Injectors
Various injector available: packed column injectors, split/splitless capillary column injectors or cool on-column injectors
Detectors
FID
Max temperature control 400oC
LOD ≤5 x 10 -12 G/s [n-C16]
Linear dynamic range ≥107
Drift ≤5 x 10 -13A/30min 
Noise ≤2 x 10 -13 A
TCD
Max using temperature 400oC
Sensitivity ≥10000mv.Ml/mg [n-C16]
Linear dynamic range ≥104
Drift ≤ 10 0uV/30min
Noise ≤20uA
FPD
Max using temperature 250oC
LOD ≤2 x 10 -13 g/s(P),≤5 x 10 -11 g/s (S)
Linear dynamic range ≥103(P),≥102(S)
Noise ≤2 x 10 -12 A
Drift ≤4 x 10 -11A/30min
ECD
Max using temperature 350oC
Drift ≤5 0uV/30min
LOD ≤1 x 10 -14 G/ml(y-666)
Linear dynamic range ≥104
Noise ≤20uA
NPD
Max using temperature 400oC
LOD ≤5 x 10 -13 g/s(P),≤1 x 10 -12 g/s (N)
Noise ≤4 x 10 -13 A
Drift ≤2.5 x 10 -12A/30min
Linear dynamic range ≥103(P),≥103(N)
PDHID
LOD ≤5ppb

·Application