Temperature range: | +25C~+40C |
---|---|
Accuracy: | +0.5,-0.5 |
Theory: | Temperature Recorder |
Usage: | Industrial |
Brand Name: | OTS JUNYUAN |
Model Number: | JY-OTS-1000P5 |
Place of Origin: | China (Mainland) |
Test Dust: | Dry Talcum Powder |
Quick Details
Specifications
IP Class Sand Dust Proof Edurance Test Chamber for Electrical Appliance
Working principle of chamber:
Centrifugal Fan is the main source of dust movement, the high-speed operation of the fan so that dust into the fan due to the centrifugal force of the Role of fan blades into the dust pipe dust. the wind speed of the Pipeline is usually more than 7.5m / s (too low wind speed will make the Dust precipitate in the pipeline will cause the test tube can not Work). the dust line was sprayed from above into the test chamber and Passed through the sample at a rate of> 2 m / s. Sand and dust Chamber As the lower part of the working box for the funnel, it can Precipitate down the dust into the lower part of the 72 50um filter. through the vibrating motor vibration, less than the diameter of the Filter hole in the dust and centrifugal fan sucked so back and forth.
the main technical parameters
Electronic and Control systems:
1. the main controller LCD touch screen, can do programmable test
2. PLC complex man-machine interface control, contrived moving time, rest time and power-on time setting.
3. Multiple sets of test procedures can be planned in the interface, and Each group can be arbitrarily set the desired action, easy to operate.
4. If a fault occurs in the operation of various signals are written records will be displayed to facilitate maintenance.
5. Touch the setting, digital direct display, all the conditions fully automatic locking by a microcomputer.
6. Imported electronic temperature and humidity sensor, not need to cleanup.
Packing & Shipping
Working principle of chamber:
Centrifugal Fan is the main source of dust movement, the high-speed operation of the fan so that dust into the fan due to the centrifugal force of the Role of fan blades into the dust pipe dust. the wind speed of the Pipeline is usually more than 7.5m / s (too low wind speed will make the Dust precipitate in the pipeline will cause the test tube can not Work). the dust line was sprayed from above into the test chamber and Passed through the sample at a rate of> 2 m / s. Sand and dust Chamber As the lower part of the working box for the funnel, it can Precipitate down the dust into the lower part of the 72 50um filter. through the vibrating motor vibration, less than the diameter of the Filter hole in the dust and centrifugal fan sucked so back and forth.
the main technical parameters
Sand and dust test chamber | |||||
Model | JY-OTS-500P | JY-OTS-800P | JY-OTS-1000P | JY-OTS-1500P | |
Workroom size(DWH)mm | 800800800 | 80010001000 | 100010001000 | 100015001000 | |
Technical index and material | Inner | SUS import stainless steel plate | |||
Shell | Cooling steel plate,surface spray | ||||
Electrical control systems | PLC Controler | ||||
Temperature control | Touch- temperature controller | ||||
Blower Time | 1S-99H(optional) | ||||
Vibration time | 1S-99H(optional) | ||||
Timing range | 0-99H | ||||
Dust dosage | 2-4kg/m | ||||
Dust dimensions | <=32m33%.32m-250m67% | ||||
Mesh Diameter | 50um | ||||
Wire Spacing | 75um | ||||
Mode of operation | Procedure | ||||
Dehumidify | Electric heating | ||||
Temperature range | 25C-40C | ||||
Fairlead | 50mm. with sealing plug | ||||
Inspection window(mm) | Visible glass window with scraper | ||||
the time of lighting and energize | Adjustable opening and closing | ||||
Pumping time | 60 times ,Gas volume inner of Hull/hour | ||||
Pressure differential | 2kPa(20mbar) | ||||
Blowing machine | Stainless centrifugal fan | ||||
Vacuum system | Pressure gauge,air filter regulator,triple pieces of Voltage,connecting pipe | ||||
Electric components | Contactor, switch and button import from France Schneider and FULI" japan | ||||
Electric socket | Dust-protected socket,16A 220V | ||||
Standards Applied | GB4208-93 method of dust proof experiment ;method of dust proof experimen; GB2423.37-89 basic environmental test of electronic product for clectrician LA. Free-falling dust; Satisfy rank of protection: IP5X .IP6X Specific parameters stander by this specification . |
1. the main controller LCD touch screen, can do programmable test
2. PLC complex man-machine interface control, contrived moving time, rest time and power-on time setting.
3. Multiple sets of test procedures can be planned in the interface, and Each group can be arbitrarily set the desired action, easy to operate.
4. If a fault occurs in the operation of various signals are written records will be displayed to facilitate maintenance.
5. Touch the setting, digital direct display, all the conditions fully automatic locking by a microcomputer.
6. Imported electronic temperature and humidity sensor, not need to cleanup.
Packing & Shipping
Packaging Details:Strong wooden case
Delivery Detail: 35 days after getting the payment
Our Services:
We are here to help you. Services available:
Replacement manuals
- Consumables and replacement parts
- on-site maintenance and repair
- Warranty support
- over-the-phone service support and advice
If you have an OTS chamber , we will do our best to help you in whatever way we can.
Manuals: PDF Files of manuals for current models are available free. Please contact Us to request a copy. Older and printed copies at additional charge.
- Service repair: We Prefer you contact us direct at the factory with any problems you may Be experiencing. We help diagnose the problem, arrange for any needed Parts, and schedule a technician to do the repairs. Rarely can a local Contractor get all that done for you as quickly as we can. We also Guarantee the work.