Product Main

Quick Details

Brand Name: NORDA
Name: particle counter
Certificate: NORDA
Weight: 7kg
Usage: testing clean rooms
Model Number: ND series
Place of Origin: China (Mainland)
Name: laser particle counter
Flow rate: 28.3PLM,50,100PLM
usage: clean rooms

Specifications

 : ND6100,  ND6101,ND6102,ND6103  


Selecting between 28.3L/min or 50L/min, or 100L/min

ND6100 laser particle counter (28.3L/min,50L/min,100L/min) is a kind of instrument used to measure the size and quantity of dust particles in unit volume air of clean environment, which can directly detect different class clean environment.

with features of using laser light source, large LCD display, small size, light weight, high precision, simple operation, storing & printing the measurement results and so on, this product is widely used in electronics, optical, chemical, food, cosmetics, e-health, biological products, aerospace, etc.


Features:  

1.Flow rate is between  28.3L/Min(1CFM),50L/min,100L/min.

2. AC DC power supply (Built-in  lithium battery).

3.Stainless steel housing, easy to surface cleaning and disinfection. 


4. Display six-channel particle sizes and particle concentration (pcs / m3),you can have a selecting.
5. Automatic UCL calculation (for the 2010 version of GMP, Euro version of GMP, ISO14644, 16292-2010,etc.)

6.95% UCL calculation of particle concentration can be displayed directly (particles / cubic meter, grain / cubic feet)

Specification:

Model ND-6100 ND-6101 ND-6102
Overall dim.(mm) 230 x 220 x 210 (Wide*Deep*High)
Display 7 inch LCD Touch screen
Flow rate 100L/min 28.3L/min 50L/min
Weight about 7.3Kg
Power supply DC24 V(Built-in Lithium Battery)
Size channels 0.5,1,3,5,10,25(m)
Optional: 0.3,0.5,1,3,5,10(m)

0.3,0.5,1,3,5,10(m)
Calculation 95% UCL calculation of particle concentration can be displayed directly (particles / cubic meter and others)
(for  ISO14644-1,the 2010 version of GMP, Euro version of GMP, 16292-2010,etc.)
Sample time 0~9999s (selectable)
Zero count 10min
Location labels 0~9999
Print Built-in Thermal Printer
Alarm 100, 1000, 10000 Class alarm
Date storage 10000  groups 
Max. Particle concentration 35000 particles/L
Light source Semiconductor laser(30000 hours mean time between failures)
Working condition Temperature:10~35;  Humidity:20%~75%RH;  Atmosphere Pressure: 86~106Kpa
Optional Trolley for clean room

laser particle counter with1CFM 283L50L100Lmodel ND series

Information,contact us~