Product Main

Quick Details

Brand Name: senshang
Material: steel
Color: silver
Packing: carton
usage: measure NO2 or NOx

Specifications

1. Introduction

 NOx gas converter is used for the application that using NO gas analyzer to measure NO2 or NOx. It can convert NO2 to NO in the air or from the stack. this is an optional equipment used before NO gas analyzer, which can got NOx (NO+NO2) value by converting NO2 to NO.

2. Features

 High conversion rate under low temperature.            

 up to 300ppm NO2 switching capacity

 A long lifespan

 the patented conversion tube structure

 Micro processing controller intelligent temperature control

 the temperature can be adjusted freely

 with the temperature alarm

 with the LED display

 19 "rack enclosure and wall-mounted type optional

 Easy for replacement of special switching tube design

3. Specifications

 Working temperature: 365            

 Prewarming time: 30 minutes            

 Sample gas pressure: 1.5bar            

 Sample gas flow rate: 120L/h            

 Sample gas temperature: 5-80            

 Dew point: <10 C            

 Environment temperature: 5-50            

 Storage and transportation temperature: -20 - 70            

 Relative humidity: <80%RH            

 Power supply: 220VAC, 50/60Hz            

 Power consumption: about 400W            

 Eack installation: 19 ", 3U or wall-mounted (different type)           

 Weight: about 5.5Kg            

 the protection grade: IP20            

 the conversion efficiency >95% (new conversion tube)            

 the conversion tube service life: more than 12 months, depending on the concentration of NO2            

 the maximum capacity: 300ppm (flow 60L/h)

NOx Converter for cems and vocs