Quick Details
Specifications
Sigle-phase/up to 7 functions
Double load for paralleled testing; 7 functions in 1 electrical safety and
Properties' test; double channel display together; electric ARC, load open-circuit
and short-circuit test function.
Applicable to such small home appliances high-speed product lines.
Technical characteristics:
Compliant with GB4706.1-2005,
IEC60335-1:2004, UL60335-1:2002, GB4943.1-2001, standard requirements.
Three phase, single phase load compatible; Three phase rated load
Capacity 20kVA, can custom 60kVA.
Seven functions into one device:
Ground bond, insulation resistance, AC hipot, DC hipot, leakage current, power
Test and low voltage start test.
15-group / 8-step test condition,
Intelligent judgment, auto compensation, auto protection;High speed test, the
Fastest speed can be 0.5s.
Voltage ramp-up and ramp-down,
Discharge function;Dynamic and static leakage current, hipot test.
Auto compensation, auto protection,
Auto memory, auto judgment, voice and light alarm.
Serial communication, print, foot
Pedal interface, outer alarm light, PLC, network card interface.
for safety test, the precision is
2%, for power test, the precision is 5%;for DUT, can support both grounding and
Suspending.
Function preview, single screen
Setting, no need flip, simple operation.
Support both 50Hz and 60Hz frequency, Chinese and English operation Interface.
Model |
AN9651TH (F)(Three Phase/single phase7 In1) |
Ground Bond Test | |
Empty Load Output Voltage |
<12V |
Output Current Setting Range |
AC3A~30A 50Hz/60Hz,1A/step |
Output Current Setting Accuracy |
(2% of setting+0.2A) |
Upper Limit of Resistance |
3Aoutput current10A, 10~600m; 10A |
Ground Bond Resolution/ Measurement Range/ Accuracy |
1m, 10~600m, (2% of reading +2m) |
Compensation Resistance Measurement Range/Accuracy |
0~200m 1m |
Test Time Setting Range |
0.5~999.9s 0.1s/step |
Insulation Resistance Test | |
Rated Output Load/Current |
1VA (1000V/1mA) 1mA |
Output Voltage Range/Adjustable |
DC:500V/1000V <1% (Empty load to full load) |
Output Voltage Ripple |
Pure resistance load, 1000V/1mA, <3% |
Insulation Resistance Range |
1.0~2000M |
Insulation Resistance Lower Limit Setting Range/ Resolution |
1.0~99.9 M, 0.1M/step, 100~2000 M 1 M/step |
Insulation Resistance Accuracy |
1~200 M: (2% of reading+2counts), 200~2000 M: (5% of reading+5counts) |
Ramp-up Time Range/Resolution |
0.1~999.9s 0.1s |
Delay Judgment Time |
0.3~10.0s (Count from ramp-up ) |
Test Time Setting Range |
0.5~999.9s 0.1s/step |
AC Hipot Test | |
Rated Output Load |
500VA (AC:5000V/100mA) |
Output Voltage Range /Frequency/Adjustable Rate |
500~5000V, 50Hz/60Hz, <2%(Empty load to full Load) |
Output Voltage Setting Range/Accuracy |
Cold state AC:500~5000V, Hot state AC:500~2500V, 5V/step, (2.5% of setting+5V) |
Leakage Current Preset Alarm |
Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step |
Compensation Current Setting Range/ Resolution |
0.00 ~5.00mA, 0.01mA |
Output Initial Voltage Range |
(0%~99%) of test voltage, from 0V to initial voltage Need 200ms |
Leakage Current Measurement Range/Resolution/Accuracy |
0.10~99.99 mA, 0.01mA, (2% of reading +2counts) |
Test Time/Ramp-up/Ramp-Down |
0.5~999.9s, 0.1~999.9s, 0.0~999.9s(after test pass) |
DC Hipot Test | |
Rated Output Load |
60VA (AC:6000V/10mA) |
Output Voltage Range /Frequency/Adjustable Rate |
500~6000V, <2%(Empty load to full load) |
Output Voltage Setting Range/Accuracy |
Cold state AC:500~6000V, Hot state AC:500~2500V, 5V/step, (2.5% of setting+5V) |
Leakage Current Preset Alarm |
Upper limit: 50~9999A; Lower limit: 0~9999A, 1A |
Compensation Current Setting Range/ Resolution |
0 ~5000A, 1A |
Output Initial Voltage Range |
(0~99%) of test voltage, from 0V to initial voltage Need 200ms |
Leakage Current Measurement Range/Resolution/Accuracy |
50~9999A, 1A, (2% of reading +2counts) |
Test Time/Ramp-up/Ramp-Down |
0.5~999.9s, 0.1~999.9s, 0.3~10.0s(Count from Ramp-up) |
Leakage Current Test | |
Output Voltage Range |
1.06 times of input voltage (output by isolated Transformer) |
Leakage Current Range/Resolution |
Upper limit: 50~9999A 1A |
Compensation Current Setting Range/ Resolution |
0~500A 1A |
Test Time Range/Resolution |
1.0 ~999.9s 0.1s |
Leakage Type |
Dynamic, static |
Three Phase Dynamic Mode |
A mode: ABC-AB-BC-CA (standard) |
Output Voltage Measurement Range/Accuracy |
AC 50~500.0V 0.1V, (0.5% of reading+2counts) |
Leakage Current Measurement Range/Accuracy |
50~9999A 1A |
Leakage Current Accuracy |
50.0~2000A is (5% of reading);2000A or above is (2.5% of reading) |
Leakage Power |
6kVA,isolated transformer capacity is designed According to the rated voltage 220V |
Time Setting Range |
1~300s |
Measurement Mode |
Dynamic, static |
Power Test | |
Output Voltage Range |
1.00 times of input voltage (output by isolated Transformer) |
Power Setting Range/Resolution |
Upper limit, Single phase:20~6000W Three
Phase:60~20000W |
Test Time Range/Resolution |
0.5 ~999.9s 0.1s |
Voltage Measurement Range and Accuracy |
AC:50.0V~500.0V, 45~65Hz: (0.5% of reading +2counts) |
Current Measurement Range and Accuracy |
AC:0.30~30.00A, 45~65Hz: (0.5% of reading +2counts) |
Active Power Measurement Range and Accuracy |
20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW |
Power Factor Range and Accuracy |
0.100~1.000, (2% of reading +2counts) |
Low Voltage Start Test | |
Output Voltage Range |
0.85 times of input voltage (output by isolated Transformer) |
Test Time Range/Resolution |
0.5 ~999.9s 0.1s |
Current Setting Range/Resolution |
Upper limit: 0.30 ~30.00A Lower limit: 0.00 ~30.00A0.01A |
Voltage Measurement Range and Accuracy |
AC:50.0V~500.0V, 45~65Hz: (0.5% of reading +2counts) |
Current Measurement Range and Accuracy |
AC:0.30~30.00A, 45~65Hz: (0.5% of reading +2counts) |
Display Mode |
10.4 Inch, 640480, TFT LCD display |
Isolated Transformer Capacity |
Three Phase:20kVA, Single Phase: 6kVA |
over Current Protection |
Power, Start, Leakage Current, Phase Current35A, over 5s protection |
DUT Type Setting |
Single Phase, Three phase Three line, Three phase Four line |
Grounding Mode |
Load Grounding, Load Floating |
DUT Error Processing |
Voice and light alarm |
Outer Alarm Light |
Red alarm light |
PLC Output Interface |
Each test function have one |
Interface Function |
RS232, Outer Printer, Remote |
Optional Interface |
RS485, GPIB, Internet Card, Inner Printer |
Remote control Interface |
Foot Pedal, three-color alarm light(on-off signal Output, PRINTER, 97 COM interface, Mini-Printer function optional) |
Test Group |
15 groups, 8 steps / group |
Buzzer Alarm Voice |
Can turn on,Can cut |