Brand Name: | Readeagle |
---|---|
Type: | instrument |
Properties: | The Basis of Surgical Instruments |
Material: | Metal |
Voltage: | 100~240V |
Place of Origin: | China (Mainland) |
Model Number: | Re902-C1 |
Instrument classification: | Class III |
Display: | 8.4"color touch screen and button operation |
I:E Rate:: | 8:1-1:8 |
Breathing Frequency: | 1-65BPM |
O2 Flush: | 25-75L/min |
Working Mode:: | Semi-open, Semi-closed and Closed |
Tidal Volume:: | 100-1600ml |
Driven Mode: | Pneumatically dirven and electronically controlled |
Respiration Mode: | IPPV, SIMV, Sigh, Manual, Demo, Standby |
Quick Details
Specifications
Technology Specification:
Patient
Suitable for Adult and Pediatric
Display
8.4"color touch screen and button operation
Optional: 15.1"color touch screen
Frame
Metal alloy frame, with plastic spray process, durable and elegant. Two big drawers, convenient for different articles storage
Flowmeter
Classical Flowmeter of 5 tubes, with O2 and N2O Linkage Device, and make sure Oxygen Concentration not lower than 25%.
O2 Flow Rate: 0-1L/min, 1-10L/min
N2O Flow Rate: 0-1L/min, 1-10L/min
Air Flow Rate: 0-10L/min
O2 Flush
25-75L/min
Circle Absorber
Classical CO2 Circle Absorber with twins canister
with non-return respiratory valves
with APL valve, the airway pressure could be adjusted freely
with " Manual / Ventilation" one switch
Canister Volume: each 1000ml (1000g soda lime); total 2000ml (2000g soda lime)
High Precise Anesthesia Vaporizer
Concentration Scope: 0.2-5VOL%
Flow Scope: 0.2-15L/min
Temperature Scope: 15-35
Anesthesia Vaporizer, automatic compensation for Temperature, Flow and Pressure.
Anesthesia Agent: Halothane, Enflurane, Isoflurane, Sevofluane for choice.
Monitoring Parameters
Tidal Volume
Respiratory Frequency
Minute Ventilation
FiO2
Airway Pressure (Ppeak, Ppeep)
Respiration mode
Monitored Waveform
P-t, F-t
Alarm Parameters
Airway Pressure high / low limit
FiO2 high/low limit
Minute Ventilation high/low limit
Breathing Frequency high/low limit
AC failure
O2 supply low
Alarm silence
Gas supply Pressure
O2 N2O Air
O2: 0.28-0.60Mpa
N2O: 0.28-0.60Mpa
Air: 0.28-0.60Mpa
Working Mode
Semi-open, Semi-closed or Closed
Driven Mode
Pneumatically driven and electronically controlled
Respiration Mode
IPPV, SIMV, Sigh, Manual, Demo, Standby
IPPV(Pressure Protection)
SIMV((Pressure Protection, Time protection)
Breathing Frequency
1-65BPM
Tidal Volume
100-1600ml
I:E Rate
8:1-1:8
Pressure Limit
Min. Pressure Limit: 1cmH2O
Max Working Pressure Limit: 60cmH2O
Airway Pressure Monitored
-10-60cmH2O
PEEP
0-20cmH2O (Mechanic)
Anesthesia Ventilator System Compliance
4ml/100Pa
Power Supply
220V, 50Hz; or 110V, 60Hz
Power: 90VA
Optional
Patient Monitor. YOKE .CO2 Monitor . AGSS