| Brand Name: | Hlilfe |
|---|---|
| Type: | Biochemical Analysis System |
| Range: | Liver Function,Kidney Function,Blood Fat And Blood Sugar,Myocardial En |
| color: | white |
| weight: | 6kg |
| Place of Origin: | China (Mainland) |
| Model Number: | 800C |
| display: | 5.7 LCD screen |
| Warranty: | 1 year |
| Item: | liver/kidney function, blood sugar, blood fat, ions,etc |
| System: | Fully close optical system |
| Wavelength: | 505, 510, 546, 578, 620, 630nm |
| Reagent: | All reagents workable for end point assay |
| Dimensions: | 490 mm(H) * 435 mm(W) * 280 mm |
| Printer: | Built-in thermal micro-printer |
| Function: | Detecting Body Health |
Quick Details
Specifications
1.Testing method: End point assay, and multi-point calibration assay, both covering all the
Biochemical test items
2.Wavelength:505, 510, 546, 578, 620, 630nm
3.Reagent: All reagents workable for end point assay
4.Equipped with international standard RS-232 serial port, can be upgraded with supporting
Software when connected to PC
5.Printer: Built-in thermal micro-printer, to print complete report, Can print real-time report and
Complete report
6.Display: Blue Screen, with backlight, 320*240 graphic dot matrix LCD
7.User interface: User-friendly, man-machine conversational operation, easy to learn and operate
8.Storage: 200 item parameters in storage, test item and parameters adjustable; Can store at
most 800 patients' data and hundreds of thousands of test records
9.Power supply: AC220V/50Hz
10.Working Environment: Temperature: 5-35; Relative humidity: 40~85%RH
11.Dimensions/Weight: 490 mm(H) * 435 mm(W) * 280 mm(D); 6 kg
Absorbance: |
0-3.0ABS |
Resolution: |
0.001 ABS |
Stability: |
<0.01 ABS |
Repeatability: |
<0.01 ABS |
Linearity: |
<1% |
Excursion: |
<0.002 A /h |
Cross-contamination: |
<2.0% |
Sample Volume: |
200-9999L (adjustable) |
the Advantages of Semi Auto Biochemistry Analyzer
1) Blue Screen, with backlight, 320*240 graphic dot matrix LCD
2) Thermal micro-printer
3) 6 kinds of wavelength:505, 510, 546, 578,620, 630nm
4) 2 types of testing methods: end point assay,multi-point calibration assay, every method covering all the biochemical test items
5) Can store 800 patients' data and hundreds of thousands of test records
6) Real-time clock display(Year, Month, Day, Hour, Minute, Second)
7) Data exchange can be realized when the analyzer is connected to computer via RS232C serial port, including adjusting test items, parameter, check and save test results
8) Automatically save test results, print English real- time report and comprehensive report
9) Names of the hospital and tester can be typed in; English name and gender can be input in comprehensive report, thus making the test result more complete
10) All the operating instructions are in English, and man-machine conversation is realized