Brand Name: | HAMILTON |
---|---|
Measurement range: | 14 |
Measurement error: | 0.5 |
Instrument resolution: | 10 |
Instrument stability: | 10 |
Model Number: | Inchtrode N75F 238346 |
Place of Origin: | Switzerland |
Quick Details
Specifications
Specifications
A-length
143 mm
Electrical Connector
VP6
Measurement Principle
Combination electrode; pH potential measured against reference
Measuring Range
PH 0 to 14
Sensitivity
57 to 59 mV / pH at 25 °C
Zero Point
0 ± 20 mV
PH Glass
Hamilton Type HF Glass: for < 0.01M / 200 mg/l HF at 20 °C or 0.05M / 1000mg/l HF at 50 °C
PH Glass Resistance
Moderate
Membrane Shape
Flat
Diaphragm
Single Pore ring
Number of Diaphragms
1
Electrolyte
Polisolve
Reference System
Hamilton Everef-L
Liquid Earth
Yes
Temperature Sensor
Pt1000
Diameter
22 mm
Process Connection
NPT " Thread
Wetted Parts
Glass
PEEK (Polyetheretherketone)
Serial Number
Yes
Recommended Storage
Hamilton Storage Solution or 3 M KCl
Certificate
Yes, with measured values in Buffer pH 4.01 and pH 7.00
ATEX Approval
II 1/2 G Ex ia IIB T4/T5/T6 Ga/GB; II 1/2 D ia IIIC T x °C Da/Db
IECEx Approval
II 1/2 G Ex ia IIB T4/T5/T6 Ga/GB; II 1/2 D ia IIIC T x °C Da/Db
Autoclavable
Yes
CIP
No
Steam Sterilizable
Yes
Operating Temperature Range
-10 to 130 °C
Sample min Conductivity
2 S/cm
Pressure Range bar g
0 to 10 bar at 25 °C, 0 to 6 bar at 130 °C