HTAJ series Ethylene oxide sterilizer is a low temperature sterilization equipment.

Main Advantages:

HTAJ-20cube series large-scale ethylene oxide sterilizer is a kind of low-temperature sterilization equipment: this product is designed by using advanced computer technology, automation technology, welding technology and process on the basis of absorbing the advantages of similar products domestic and overseas.

 

Technical details:

[Air replacement] Automatic ventilation, the number and interval of ventilation can be set freely.


[Full-automatic control system] One-touch start to finish, the whole process is automatically completed with printing records.

[Display status] Display temperature, humidity, pressure, disinfection time, residual gas treatment time and frequency.

[Printer] Record and print the disinfection information for the whole process for verification.

[Humidity sensor] the built-in humidity sensor automatically detects the humidity in the cavity to ensure that the humidity in the cavity is not less than 45% RH.

[Pressure sensor] the built-in pressure sensor automatically detects the pressure in the cavity to ensure effective disinfection of the pressure in the cavity.

[Temperature sensor] the built-in temperature sensor automatically detects the temperature in the cavity to ensure that the temperature in the cavity is maintained within the disinfection range of 50 ° C ± 3 ° C.

[Humidification system] Pulsed humidification under vacuum conditions, to ensure that the humidity in the cavity is uniform and controllable; the humidity in the disinfection room should be controlled within the range of relative humidity 45% -80%.

[Residual gas treatment device] the removal rate of the filter with a particle diameter ≥ 0.3 μm is not less than 99.5%.

[Leak rate in sterilization room] the leakage rate is measured under the condition of no-load and sterilization room temperature, the test pressure is -50kPa, and the measurement time is 1h.

[Vacuum speed] under no-load conditions, the time for the disinfection cabinet to reach -60kPa from normal pressure should be ≤30min.

[Vacuum system] the oil-free vacuum pump is used, which is simple and convenient, the vacuum rate is fast, and the service life is long.

[Heating system] It adopts explosion-proof electric heating device, which is more convenient and quick.

[Tray rack] Use 304 stainless steel tray rack.

[Various protection functions] with over-voltage and over-temperature protection function, it is more safe and reliable.

[Casters] Yes.

[Alarm system] There is a high temperature alarm system and an ultra-low pressure alarm system; when the disinfection temperature of the disinfection room exceeds 59 ° C, it should be able to sound an alarm.

[Category] Class II medical instruments.