Brand Name: | CEAULIGHT |
---|---|
usage: | photolysis of water |
Weight: | 25kg |
Temperature range: | -20-80 |
Power supply: | 300W |
Model Number: | CEL-PCRD300-12 |
Place of Origin: | China (Mainland) |
Quick Details
Specifications
Instrument advantages
1) Use side irradiation to increase the irradiation area, which is 10 times that of bottom or top irradiation;
2) 12-bit independent numerical control, stirring, light intensity, multi-wavelength, ventilation, vacuuming;
3) the wavelength can be arbitrarily matched; optional wavelengths 365nm, 395nm, 405nm, 420nm, 455nm, 470nm, 500nm, 520nm, 590nm, 620nm, 660nm, 740nm, 810nm, 850nm, 940nm, white LED;
4) Simultaneous stirring of the 12-bit reactor is realized to better mix samples;
5) Modular design can be used to achieve multi-band irradiation only by replacing the PCRS25 module;
6) the LED light source does not need water cooling, no filter, and the light is uniform;
7) the LED power supply is individually controlled by the touch screen to achieve free matching.
Technical Parameters
items |
Parameters |
Function
|
12-bit photochemical reaction instrument; independent control, stirring, adjustable light intensity, multi-wavelength selection, ventilation; PCRS25 photochemical reaction instrument light box realizes online hot swapping and replacement of band. |
LED Standard 12-bit wavelength |
365nm, 395nm, 405nm, 420nm, 455nm, 470nm, 500nm, 520nm, 590nm, 620nm, 660nm, white light |
Optional wavelength |
740nm, 810nm, 850nm, 940nm |
Control mode |
7-inch high-definition touch screen CNC, 240 * 128dpi |
Input power |
300W,AC110-240V,±0.01% |
LED power consumption |
12*25W, DC38V,±0.01% |
Irradiation
|
12-bit independent LED side irradiation, irradiation area 100mm * 20mm (50ml); optional 100mm * 30mm (100ml) |
Number of samples
|
12-bit 50ml quartz reaction tubes + temperature controlled cold well (optional 100ml) |
Irradiation unevenness
|
5% |
Standard configuration
|
1 set of host system, 1 set of CEL-GDT-12 gas distributor, 12 sets of CEL-PCRS25 photochemical reactor light box, 14 sets of quartz reaction tubes (50ml), 12 sets of PEEK special joints and pipelines |
Description of wavelength selection: customers can choose the band and quantity according to the experimental requirements, a total of 12 standard; can also add multiple bands according to demand. |