Product Main

Quick Details

External Size: Export standard packing
Type: Tuning Fork Density Meter
Protection Level: IP67
Brand Name: NUOYINGJIAYE
Color: blue
Model Number: NYDE
Place of Origin: China (Mainland)

Specifications

NYDE Tuning Fork Resonance Density Meter 4~20Ma Or RS485 Modbus RTU 24VDC

NYDE Tuning Fork Resonance Density Meter

 

the NYDE plug-in density meter adopts plug-in installation and is widely used in pipelines. the density of medium in open tanks and closed tanks is directly determined by the vibration frequency received by the tuning fork of the sensor inserted into the medium. the built-in temperature sensor of the sensor provides temperature compensation for it.

Scope of application: the plug-in density meter can perform density detection online in real time. It can be used for process control of products with density as the basic parameter or quality control with solid percentage or concentration percentage as a reference. Typical industries include petrochemical industry, wine industry, food industry, pharmaceutical industry and mineral processing (such as clay, carbonate, silicate, etc.), which are specifically used in the interface detection and mixing of multi-product pipelines in the above industries the density detection, the end-point monitoring of the reactor, and the interface detection of the separator.

 Modbus RTU Tuning Fork Density Meter Resonance Liquid Density Sensor

NYDE Tuning Fork Resonance Density Meter Technical Parameters

 

Measuring range 0 2.5 g /cc (0 2500 kg/m3),0~100.0%
Calibration range 0 2.5 g /cc (0 2500 kg/m3),0~100.0%
Measurement accuracy ± 0.002 g /cc (± 2 kg/m3),±0.5%
Repeatability ± 0.001 g /cc (± 0.2 kg/m3) ±0.1%
Medium temperature range -50℃~+150℃
Working pressure 4mpa
Fluid viscosity range 0600 cP
Temperature Coefficient 0.1 kg/m3/℃(±0.5%) after correction
Stress effect Negligible
Built-in temperature sensor Digital sensor
Wetted material 316L,2205 Stainless steel,HAC,Titanium alloy
Fork body coating Standard PTFE or PFA
Power supply 24VDC,≥50 mA
Analog signal output 4-20 mA,0-1000Hz,RS485 Modbus RTU
Output accuracy (20℃) ±0.1% of reading or ± 0.05% FS
Output repeatability (-40~+85℃) ± 0.05% FS
Process connection DIN 50 PN16 DIN 50 PN40
IDF and RJT sanitary
Protection level IP67
Shell Aluminum alloy